GENERAL CLEANING

 

After each medical procedure, the table should be properly cleaned.  Do not use harsh 
abrasives, solvents, sprays, or corrosive agents.  Some accessories may come with 
individual cleaning instructions. 

APPROVED AND TESTED DISINFECTANT CLEANERS FOR TABLE: 

 

Sodium hypochlorite (generic household bleach) in solution of 5.25% sodium 
hypochlorite diluted between 1:10 and 1:100 with water. 

 

Alcohol (generic). 

 

Envirocide ®

 

Disinfectant and Cleaner. 

APPROVED AND TESTED DISINFECTANT CLEANER FOR RADIATION SHIELD: 

 

Scrubbles

 (Infab Corporation) 

  https://www.infabcorp.com/apron-cleaning/

 

APPROVED AND TESTED GENERAL PURPOSE CLEANERS: 

 

Simple Green ™ cleaner. 

CLEANING STEPS FOR TABLE: 

a)

 

Move the tabletop to a level horizontal position. 

b)

 

Lower the tabletop to its lowest position. 

c)

 

Disconnect the table from the AC power outlet and press the Emergency Stop 
Pushbutton. 

d)

 

Power cord, Handset Control and Foot Control cords must be plugged in at the table 
base to protect the inside of the connectors from debris. 

e)

 

Remove all pads and accessories. 

f)

 

Wipe off any excess fluids with a water dampened cloth or sponge. 

g)

 

Clean the Table Top and accessories using an approved cleaner listed above. 

h)

 

Clean all pads according to the instructions attached to the pad. 

i)

 

Clean the table frame, castors, and base with Simple Green ™ cleaner. 

j)

 

Thoroughly rinse Patient Mattress Pad, Table Top and Accessory Rails with water. 

k)

 

Gently rub with a soft clean cloth until dry. 

 

CLEANING STEPS FOR RADIATION SHIELD: 

a)

 

 Lift off the upper section and lay it flat before using the recommended cleaner in an 
adequately ventilated area. 

b)

 

Apply approved cleaner to one side at a time and allow to stand a few minutes. 

c)

 

Scrub with a soft bristle scrub brush.  Do not let the solution dry before rinsing. 

d)

 

Rinse with water and a damp cloth. 

e)

 

Scrub and rinse again, if necessary. 

f)

 

Remove the main section of the Radiation Shield and clean in the same manner. 

g)

 

Attach Radiation Shield back onto Tabletop. 

h)

 

Install upper panel of Radiation Shield. 

 

Содержание Aspect 100-4 Q-CARD

Страница 1: ...Made in the USA with global materials Aspect 100 4 Mobile Image Table Operator Manual...

Страница 2: ...agencies regarding specific requirements and regulations applicable to the use of this equipment Image Diagnostics Inc certifies only the equipment After sale operating practices and safety are the re...

Страница 3: ...3 TABLE MOTION CONTROLS 15 9 STANDARD ACCESSORIES 16 Please reference website for most up to date list of available accessories for purchase No IDI Accessories nor components of accessories listed be...

Страница 4: ...gnostics com 4 11 GENERAL CLEANING 22 12 MAINTENANCE SERVICE REPAIR 23 12 1 RECOMMENDED PERIODIC PERFORMANCE CHECKS 23 12 2 SERVICE REPAIR STATEMENT 23 13 TROUBLESHOOTING 24 14 DISPOSAL OF COMPONENTS...

Страница 5: ...esult in death personal injury or damage to the equipment Emergency Stop Pushbutton Electric Shock hazard Information or instructions shown near this symbol must be adhered to in order to prevent a po...

Страница 6: ...d Representative Model of Table Serial Number of Table Indication of European Conformity for sale in the European Economic Area EEA Item complying to Type B applied part per IEC 60601 1 Date of manufa...

Страница 7: ...disrupted 3 SAFETY INSTRUCTIONS All persons using this equipment must fully understand its operation instructions emergency procedures capabilities including total range of motion and be aware of all...

Страница 8: ...not been tested for use with high frequency surgical equipment cardiac defibrillators or cardiac defibrillator monitors Use with such equipment may cause patient burns explosion hazards or electrical...

Страница 9: ...the safety of patient and operator CAUTION During operation of the table if any unusual sounds and or erratic movement is observed immediately discontinue use of the table CAUTION Do not use table fo...

Страница 10: ...no closer than 30 cm 12 inches to any part of the Aspect 100 4 including cables connected to the table Otherwise degradation of the performance of this equipment could result NOTE The emissions charac...

Страница 11: ...with the controls before using them External electrical power connection and disconnection is through the AC power cord and outlet The table will operate on 110 V 230 V or on internal battery backup...

Страница 12: ...assive magnetic brake it remains engaged energized in the event of a complete loss of electrical power Power is necessary to release the brakes As the braking system utilizes permanent magnetic brakes...

Страница 13: ...break and then transfer the patient onto the Table 7 4 Preparation for Performing CPR Return table to level position retract tabletop to minimize overhang and lower table position to a comfortable hei...

Страница 14: ...ully engaged The Rear Casters will be totally locked and will not roll in any direction Always lock all casters in this way before loading the patient The table is in the UNLOCK setting when the Caste...

Страница 15: ...ngaged 8 3 TABLE MOTION CONTROLS Aspect 100 4 tables are equipped with tabletop motion control boxes that allow users to move the tabletop in many different directions Continually pressing down on the...

Страница 16: ...Accessory Rail Then the Unit can be slid along the rail or removed 9 STANDARD ACCESSORIES Please reference website for most up to date list of available accessories for purchase No IDI Accessories nor...

Страница 17: ...bletop hangs down between the rails and the tabletop and wrapped over the outside of the rails to meet over the top of the table If the strap is to be used along the rest of the tabletop slide the str...

Страница 18: ...ble 6 length European version with 10mm x 25mm metric rail is available 10 2 Vascular Access Arm Board VAB One piece extra wide carbon fiber arm board for vascular access procedures Mounts to table by...

Страница 19: ...opean version for 10mm x 25mm metric rail size is available 10 4 Arm Board Shoulder Mount Carbon fiber base Mounts to table by sliding under Patient Mattress Pad Includes padded arm board and restrain...

Страница 20: ...pivoting the extension up or down to angle the patient s head To remove the Articulating Headrest Extension from the Tabletop depress the release levers located on the underside of the extension and p...

Страница 21: ...l and the tabletop by clamp 10 9 Anesthesia Screen Holder May require a pair of side rail clamps Flexible Hoop If this is to be used in an area of the tabletop that has no side rails then a pair of Cl...

Страница 22: ...cord Handset Control and Foot Control cords must be plugged in at the table base to protect the inside of the connectors from debris e Remove all pads and accessories f Wipe off any excess fluids wit...

Страница 23: ...erviceable Replacement must be performed only by a qualified service technician Weekly Check battery operation by disconnecting the AC power and running the tabletop up and down Run the table top thro...

Страница 24: ...of time 2 Contamination on brake surfaces or improper cleaning or maintenance performed 1 Release brakes using Tabletop Motion Control Panhandle and move tabletop in all directions Working tabletop mo...

Страница 25: ...Board Housing Cable Electronic Plastic Copper Table Base Frame Casters Covers Metal Steel Plastic and Steel Stainless Steel Electronic waste and batteries Electronic components and devices must be dis...

Страница 26: ...in Patient capacity 272 kg 600 lbs without tabletop extensions Emergency Stop Pushbutton Multi caster 2 Position Multi lock system Total Lock Steer Backup battery power 16 SPECIFICATIONS Mode of Oper...

Страница 27: ...equirements of FDA CFR Title 21 Chapter 1 Subchapter J Guidance and Manufacturer s Declaration Emissions All Equipment and Systems Table 1 The Aspect 100 4 is intended for use in the electromagnetic e...

Страница 28: ...1kV 2 kV Line to Ground Main power quality should be that of a typical commercial or hospital environment Voltage Dips Dropout EN IEC 61000 4 11 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT...

Страница 29: ...and between 0 15MHz and 80MHz 80 AM at 1kHz see table 5 of IEC 60601 1 2 3V m 80MHz to 2 7GHz 80 AM at 1kHz AC Mains 3V 0 15MHz 80MHz 6V in ISM band between 0 15MHz and 80MHz 80 AM at 1KHz see table 5...

Страница 30: ...tput power of the Communications Equipment Table 4 Max Output Power Watts Separation m 150kHz to 80MHz D 3 5 3 Sqrt P Separation m 80 to 800MHz D 3 5 3 Sqrt P Separation m 800MHz to 2 5GHz D 7 3 Sqrt...

Отзывы: