Installation and Maintenance Guide for IDI 1000F Series Suspension System for use
with OTC Adapter
D000-571 REV E
MAY-2018
Image Diagnostics, Inc.
Page
21
7.
Check the pad attached to the underside of the monitor mount. If there is any
damage to the pad or if the pad is not properly attached to the unit, it must be replaced.
The procedures mentioned in these instructions are based on the recommendations of
HHS, I.E.C., etc.
In cases of increased wear or where severe working conditions exist, maintenance
checks should be made at shorter than the specified intervals.
Should any control or indicator fail to operate properly, do not use the equipment until
a repair has been affected. Operating equipment with defective components may
expose the operator or the patient to safety hazards.
The manufacturer strongly recommends that a maintenance program be initiated and
that a maintenance record be kept detailing dates and nature of maintenance
performed, the name of the service engineer, and any other relevant information.
WARNING!
Removal of any hardware requires
re-assembly with Loctite #242 or
equivalent (depending on hardware
size)
following
manufacturer's
instructions.
Pad