Installation and Maintenance Guide for IDI 1000F Series Suspension System for use
with OTC Adapter
D000-571 REV E
MAY-2018
Image Diagnostics, Inc.
Page
12
ELECTRICAL CONNECTIONS
1.
Colors of conductors in power supply cords shall be in accordance with IEC
publication 60227 (amendment #1) or with IEC publication 60245.
Wiring Color Codes:
International: brown (line), blue (neutral), green/yellow (ground)
North America: black (line), white (neutral), green/yellow (ground)
2.
Cable tie mounting platforms are supplied near the terminal block for anchoring
the monitor cords and the wires from the mains. Use nylon cable ties. (Tying
the cord into a knot or tying the ends with string shall not be used for cord
anchorage).
3.
Conductors of the power supply cord shall be arranged that the protective earth
conductor is not subject to strain as long as the phase conductors are in contact
with their terminals.
4.
All power cords used for mains connection shall have double insulation.
Conduit provided for routing cables shall not be relied on for insulation.
5.
Connect all monitor ground wires and the service ground wire as shown in the
figure above. (The number of wires varies with the number of monitors
installed).
INSTALL STATIONARY CABLE HANDLING ASSEMBLY
(Refer to Stationary Cable Handling Kit Drawing)
1.
Determine which rail end the
stationary
cable handling kit will be positioned.
If, at the far end of the rail, slide kit into rail track; slide kit into rail track
prior
to positioning the
mobile
cable handling kit. If stationary kit is to be
positioned at the near end of rail, slide into rail
after
positioning mobile cable
handling kit. Stationary kit is typically mounted approximately 6 inches
(15cm) from rail end.
2.
After applying LOCTITE #242 to threads, tighten the setscrew (item 4) to
secure in place.
3.
Remove retainer (item 2) by taking out the two mounting screws (item 3).
(#10-32 HARDWARE)
NUT "B"
MAIN GROUND WIRE
FROM SERVICE
NUT "A"
under split lockwasher and nut "A".
First attach main ground wire
Attach all other ground wires
between nut "A", lockwashers,
and nut "B".