idi 1000F Series Скачать руководство пользователя страница 10

Installation and Maintenance Guide for IDI 1000F Series Suspension System for use 

with OTC Adapter

 

D000-571 REV E 

MAY-2018  

         Image Diagnostics, Inc. 

Page 

10

 

 
3. 

Take the thicker thrust ring (item 20) and put around rotation bearing (item 50). 

 
4. 

Place the stainless-steel thrust ring (item 47) below the thicker thrust ring (item 
20). 

 
5. 

The six bolts (item 16) with lockwashers (item 29) go through rotation cap (item 
43)  and  rotation  bearing  (item  50)  into  the  pivot  tube  (item  43  in  the  Core 
Assembly Drawing).  Use 

LOCTITE #242

 on hardware. 

 
6. 

Apply a very little amount of 

LOCTITE #242

 to and insert the six setscrews 

(item 24) into the six threaded holes on second rotation cap (item 43).  Thread 
the setscrews in until they contact the rotation cap (item 43).  The setscrews 
may be tightened or loosened later to adjust rotational resistance. 

 
 

WARNING! 

Failure to securely attach hardware 
may  result  in  serious  injury  to 
patient,  user,  and  damage  to 
equipment. 

 
INSTALL MONITOR  
(Refer to Monitor Mount Assembly Drawing)  

 

 

 
Maximum weight of any individual monitor. 
 

 
 
Maximum total weight of monitors.

 

 
 

 
1. 

To attach a monitor to the system, first remove a monitor mounting bracket 
(item #30) by removing its mounting screw (item #21).  Fasten the bracket to 
the monitor (hardware not provided).  Reinstall the bracket with the monitor. 

 
 

WARNING! 

Failure to securely attach hardware 
may  result  in  serious  injury  to 
patient,  user,  and  damage  to 
equipment. 

 
2. 

Monitors can be tilted and rotated. 

 
3. 

If monitor tilt resistance is inadequate and it fails to stay in the desired position, 
tighten bolt (item 21). 

25 LBS

11 kg

44 kg

96 LBS

Содержание 1000F Series

Страница 1: ...Image Diagnostics Inc Page 1 idi Manufacturer of Quality Medical Products Image Diagnostics Inc IDI 1000F Series Monitor Suspension System Model IDI 1000F 1 thru 1000F 6 With OTC Adapter Installation...

Страница 2: ...of this manual was originally written approved and published by the manufacturer in English The information in this manual is subject to change without notice CONTENTS SECTION TITLE A INTRODUCTION PRO...

Страница 3: ...el mounting and positioning The mobile package includes clip mounted bridge rails PRODUCT DATA Package Includes Mobile gantry and counterbalanced swing arm with 42 sweep radius 25 vertical travel and...

Страница 4: ...g documents Failure to follow these instructions can cause accidents resulting in serious injury to patient user and damage to equipment Caution risk of electrical shock Protective Ground This is the...

Страница 5: ...with state federal and local recommendations concerning electrical safety when used in medical installations All electrical connections shall be done by licensed approved electrician per national ele...

Страница 6: ...upport Image Diagnostics Inc 310 Authority Drive FITCHBURG MA 01420 USA Or call IDI at 978 829 0009 or send fax to 978 829 0027 Or toll free 877 304 5434 SAFETY HAZARDS ALERTS Alert Circumstances for...

Страница 7: ...em 9 and two rail spacers item 14 up to the adapter carriages item 1 and hold in place Place a pair of rail clips item 15 around the rail flanges locating rail clips under the mounting holes in the ad...

Страница 8: ...our assemblies in until the wheels contact the OTC rails then tighten the nut plate screws Use Loctite 242 on hardware INSTALL RAIL END STOP Refer to End Stop Assembly Drawing NOTE End plates have bee...

Страница 9: ...allet up and into the rails Adjust center wheels if necessary per figure on page 8 Make sure all carriage wheels are engaged onto rails before lowering Hi Jack Remove pallet and shipping bars from sup...

Страница 10: ...s in until they contact the rotation cap item 43 The setscrews may be tightened or loosened later to adjust rotational resistance WARNING Failure to securely attach hardware may result in serious inju...

Страница 11: ...l strip item 31 has been provided for distribution of line voltage to each monitor Cut and strip each monitor power cord to fit WARNING Remove plastic cover and remove block of wood between gantry cor...

Страница 12: ...nection shall have double insulation Conduit provided for routing cables shall not be relied on for insulation 5 Connect all monitor ground wires and the service ground wire as shown in the figure abo...

Страница 13: ...nut plates item 4 of two hanger standoff and safety bracket assemblies into rail track After applying LOCTITE 242 to the end that will be secured into the nutplate position standoffs approximately 48...

Страница 14: ...pounds 4 2 6 0 Nm Use LOCTITE 242 on hardware CAUTION Over tightening the standoff will weaken or break the set screw connecting the standoff to the nut plate causing attached components to fall DO NO...

Страница 15: ...d is to be attached to pivot tube and to be routed underneath the arm item 51 via the cable clamps items 46 50 Rotate entire system and bring to maximum positions the system should not drift from any...

Страница 16: ...izontal Fine adjustment to compensate monitor counterbalance can be made by turning adjustment screw item 53 Rotate adjustment nut clockwise to increase capacity counter clockwise to decrease capacity...

Страница 17: ...s with a clean cloth dampened with disinfectant or a mild detergent solution Do not use abrasives solvents sprays or corrosive cleaning agents Gently rub with a clean soft cloth to dry If room is to b...

Страница 18: ...sign reading DO NOT USE so that the system is not used by mistake Then contact your Toshiba representative for inspection and repair 2 Visual check Before checking the operation confirm the following...

Страница 19: ...ion does not lack smoothness compared to when the monitor suspension was installed There is no play There are no abnormal sounds The monitors or monitor suspension are not tilted due to looseness of s...

Страница 20: ...ete range Raise and lower monitor mount assembly Roll carriages through complete range of rails 5 Clean and touch up the painted surfaces 6 Check that the hardware is tight and that there is no wear a...

Страница 21: ...st maintenance checks should be made at shorter than the specified intervals Should any control or indicator fail to operate properly do not use the equipment until a repair has been affected Operatin...

Страница 22: ...ly recycled The gas springs are nearly 100 recyclable They must be decompressed with the remaining oil drained before processing The compressed gas inside is nitrogen J APPROVALS SGS Testing Service T...

Страница 23: ...Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with OTC Adapter D000 571 REV E MAY 2018 Image Diagnostics Inc Page 23 E May 2018 ECO E571 Added warnings for pad...

Отзывы: