Installation and Maintenance Guide for IDI Suspension System mounted to Canon DST-1000A Rails
L240-1429 REV A Image Diagnostics, Inc.
Page
17
ELECTRICAL CONNECTIONS
1.
Colors of conductors in power supply cords must be in accordance with IEC publication 60227
(amendment #1) or with IEC publication 60245.
Wiring Color Codes:
International: brown (line), blue (neutral), green/yellow (ground)
North America: black (line), white (neutral), green/yellow (ground)
2.
Cable tie mounting platforms are supplied near the terminal block for anchoring the monitor
cords and the wires from the mains. Use nylon cable ties. (Tying the cord into a knot or tying
the ends with string shall not be used for cord anchorage).
3.
Arrange the conductors of the power supply cord so that the protective earth conductor is not
subject to strain as long as the phase conductors are in contact with their terminals.
4.
All power cords used for mains connection must have double insulation. Conduit provided for
routing cables must not be relied on for insulation.
Figure 8
5.
Connect all Monitor and service ground wires as shown in Figure 10. (The number of wires
varies with the number of monitors installed).
(#10-32 HARDWARE)
NUT "B"
MAIN GROUND WIRE
FROM SERVICE
NUT "A"
under split lockwasher and nut "A".
First attach main ground wire
Attach all other ground wires
between nut "A", lockwashers,
and nut "B".