Installation and Maintenance Guide for IDI Suspension System mounted to Canon DST-1000A Rails
L240-1429 REV A Image Diagnostics, Inc.
Page
14
INSTALL MONITOR(S)
Maximum weight of any individual monitor.
Maximum total weight of monitors.
1.
On monitor mounts designed for two monitors (1000F-2), the plastic covers shown in Figure
7 will need to be temporarily removed to gain access to the rear of the monitor mounting
brackets.
2.
For each monitor to be installed, insert screws into the top two mounting holes leaving a gap
below the head of the screw big enough to fit over the thickness of the mounting bracket to
which it is being mounted.
3.
Install the monitor onto the mounting bracket by lowering the top mounting screws into the
slots on the top of the mounting bracket. Insert bottom mounting screws through the rear of
the mounting bracket into the monitor and tighten all four mounting screws. See Figure 7.
Figure 5
WARNING!
Failure to securely attach hardware may result
in serious injury to patient, user, and damage
to equipment.
25 LBS
11 kg
44 kg
96 LBS