G-4
Question
Answer
How do I prepare samples for use on the
SNAPshot Dx Analyzer?
The SNAPshot Dx work flow includes detailed
sample preparation instructions that are
based on the test information you have
selected.
In addition to the on-screen instructions,
you can reference the SNAPshot Dx sample
preparation poster.
Can I run non-bar-coded SNAP tests on the
analyzer?
Only SNAP tests that have a bar code and
appear on the SNAPshot Dx test type menu
can be run on the analyzer.
We recommend that you work through your
non-bar-coded inventory using the visual-
read work flow and the IDEXX VetLab Station
feature for recording SNAP test results.
Do I still need to warm up the SNAP device
before I run it on the analyzer?
Yes, the SNAP device must be warmed up
according to the specifications recommended
in the product insert.
Is it critical to remove the sample tube from
the incubator well after exactly 5 minutes?
Yes, the tube needs to be removed
immediately. Results may be impacted if the
sample under or over incubates.