Preparing the installation site
• Before unpacking, find a suitable space for the analyzer. Check the dimensions of the
analyzer and combine it with the needed workspace for sample handling to meet the
requirements.
• The analyzer should only be used indoors. Keep out of direct sunlight. Intense light can
interfere with the optical sensors.
• Make sure there is enough room behind the analyzer for easy connection and disconnection
of power supply, for the peripherals, and for the proper ventilation of the analyzer. The
recommended safety clearance is 10 centimeters (3.94 inches).
• Make sure you set up and operate the analyzer on a solid level surface in an environment
with fairly constant temperature and humidity. Do not operate the analyzer in temperatures
below 15°C (59°F) or above 40°C (104°F), or outside of the 20%–80% relative humidity level
(noncondensing) at 30°C (86°F). Do not expose the analyzer to strong light.
• Do not operate the analyzer in close proximity to sources of intense electromagnetic radiation
(such as unshielded intentional radio frequency (RF) sources). The analyzer is certified to
meet the electromagnetic compatibility (EMC) requirements of EN 61326-1:2007 and
EN 61236-2-6:2007.
• Do not set up and operate the analyzer in an environment with vibration sources, as these
vibrations could influence the results and/or cause instrument failure..
• Make sure the instrument is allowed to acclimatize to room temperature before use.
Installing the SediVue Dx analyzer
The SediVue Dx* analyzer works in conjunction with the IDEXX VetLab* Station.
1. Before you unpack the analyzer, choose an optimum location for the instrument. The analyzer
should be placed on a level surface in a well-ventilated area away from obvious sources of
heat, direct sunlight, cold, humidity, or vibrations.
IMPORTANT:
Ensure proper ventilation. The analyzer’s cooling vents are in the back.
2. Connect the analyzer to a numbered port on the router using an Ethernet cable (provided).
Note:
For more information about connecting your analyzer to the router, see the installation
instructions that accompanied your router.
3. Power on the SediVue Dx analyzer. Once the
SediVue Dx
icon displays on the IDEXX VetLab
Station Home screen, your connections are complete.
Note:
If the SediVue Dx icon does not appear on the IDEXX VetLab Station Home screen
within 3 minutes, contact IDEXX Technical Support for assistance.