19
2. Foldable and portable frame:
•
Lower the control unit and then fully attach the lift frame.
•
Remove the screw before folding the frame.
•
Release and support the plate before raising the frame
•
The scheme of folding the frame is shown in figure 14.
CARE AND MAINTENANCE
Failure to care for and monitor your product may result in the risk of serious injury.
Care
Clean the seats with a damp cloth. Stains can be removed with a cloth or a soft brush. Wash with a damp cloth (in clean
water) and leave to dry. Before washing, follow the instructions on the headgear label.
Do not use solvents, toilet cleaners, sharp brushes and hard objects, corrosive cleaners.
The straps can be machine washed at 30 - C. Do not use chlorine bleach.
Do not tumble dry.
Do not iron.
Maintenance
Routine inspection listed below shall be performed every day for the sake of safety.
1. Check if there is any damage or wear by visual inspection.
2. Make sure screws and nuts on lifting arm are secured.
3. Check if hook on lifting arm works properly.
4. Controller, actuator and base width adjustment verification.
5. Emergency descend function verification (both manually and electrically)
If needed, to clean the patient lift with warm water or alcohol. Make sure there is no dust and hair on the wheels.
Attention: to clean the patient lift with chemicals with Phenol and chlorine is prohibited, which may cause damage on
plastic powder or plastic material.
Inspection
Make a full inspection at least once a year. Pay close attention on those worn part such as connection and screw bolt.
Troubleshooting
Issue description
Troubleshooting