background image

Form Number TM61096 

Rev 2 Nov 2010 

IDEAL INDUSTRIES INC. 

TECHNICAL MANUAL 

MODEL 61-096 Split Jaw Smart Meter 

The Service Information provides the following information : 

Precautions and safety information

Specifications

Basic maintenance (cleaning, replacing the battery and fuses)

Performance test procedures

Calibration and calibration adjustment procedures

Form Number:    TM61096   
Revision: 2.    Date: Nov 2010

99 Washington Street 
Melrose, MA 02176 

Phone

 781-665-

1400

Toll Free 1-800-517-8431

Visit us at www.T

estEquipmentDepot.com

Содержание SPLIT-JAW 61-096

Страница 1: ...mation Precautions and safety information Specifications Basic maintenance cleaning replacing the battery and fuses Performance test procedures Calibration and calibration adjustment procedures Form N...

Страница 2: ...ification 4 Current Specification 4 Physical and environment characteristics 5 Certification and compliance 5 Required Equipment 6 Basic Maintenance 7 Opening the Meter Case 7 Replacing the Battery 7...

Страница 3: ...mation Use the Meter only as described in the Service Manual If you do not do so the protection provided by the Meter may be impaired Read the Safety Information page before servicing this product In...

Страница 4: ...es that could result in personal injury or loss of life Specific precautions Do not operate without covers To avoid personal injury do not apply any voltage or current to the product without covers in...

Страница 5: ...ption Display count 6000 Numeric update rate 5 times sec Polarity display Automatic Overrange display OL is display Low voltage indicator is indicated Automatic power off time Automatic backslit off 3...

Страница 6: ...r input voltage up to 30V Impedance increase with input voltage to 330K at 1000V Resistance Continuity Function Range Accuracy 0 0 99 9 2 reading 1 100 2000 2 reading 2 digits Overload protection 750V...

Страница 7: ...ategory III 1000 Volts Over voltage category CAT III Distribution level mains fixed installation CAT II Local level mains appliances portable equipment CAT I Signal level special equipment or parts of...

Страница 8: ...should be performed only by qualified personnel Table B Required Equipment Equipment Required Characteristics Recommended Model Calibrator AC Voltage Range 0 750V AC Accuracy 0 07 Basic Frequency Rang...

Страница 9: ...pen circuit connections are exposed To open the Meter case and do the following 1 Disconnect test leads from any live source push the button to OFF and remove the test leads from the front terminals 2...

Страница 10: ...dicated If the Meter fails any of these tests it needs calibration adjustment or repair Testing the voltage Function To verify accuracy in the AC and DC voltage ranges do the following 1 Push the butt...

Страница 11: ...esistance Function To verify the accuracy of the resistance function do the following 1 Connect the calibrator to V and COM on the Meter 2 Push the button to start your Meter 3 Apply the inputs for st...

Страница 12: ...to start your meter 3 Apply the inputs for steps 1 4 in Table 4 4 For each input compare the readings on the Meter display to the reading in Table 4 5 If the display reading falls outside of the range...

Страница 13: ...potentiometer VR13 until display reads 1005 1006 3 Set the output of calibrator for resistance 100 and connect to V and COM test probes 4 Using a small flat tip screwdriver adjust the potentiometer V...

Страница 14: ...Form Number TM61096 Rev 2 Nov 2010 Figure 1 61 096 Calibration Adjustment Points 12 VR13 VR12 VR15 VR17 VR16 VR14...

Страница 15: ...Form Number TM61096 Rev 2 Nov 2010 Figure 2 61 096 Calibration Adjustment Points 13 VR1...

Отзывы: