13
dBm (typical):
-15dBm ~ 55dBm (0dBm = 1mW into
600
Ω
)
dBm (typical):
-80dBV ~ 50dBV = 1VRMS)
Input Impedance:
10M
Ω
, <100pF
Overload Protection: 1000VDC, 750VRMS
AC Conversion Type: AC Coupled, True RMS responding
AC+DC Volts:
Same as 1.00% + 8 digits
Crest Factor:
+1.5% addition error for C.F. from 1.4 ~ 3
+3.0% addition error for C.F. from 3 ~ 4
To Measure Voltage:
1. Plug the Black test lead into the COM port and the Red test
lead into the V
Ω
port.
2. Set the rotary switch to either the V or mV position.
3. Push the blue button to choose between DC, AC, or AC+DC
measurements.
4. Connect the meter in parallel with the load or circuit.
5. Measure the voltage
6. The AC and AC+DC measurements provide a True RMS
measurement.
7. For AC Measurements, the frequency and period of the mea-
sured signal are displayed on the secondary displays.