![IDEAL 61-797 Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/ideal/61-797/61-797_instruction-manual_3692555007.webp)
13
14
Read the stored value of PI test under RECALL mode
Auto Power Off (Battery Saver)
Restore power by switching the Function Selector or by pressing
any button.
Auto Backlight
The backlight is automatically turned on in a dark environment.
Buzzer
The Meter beeps once for every valid key-press, and beeps twice for
every invalid key-press.
Ω
Ω
V
INSULATION
PASS
RECALL
PI/DAR
TEST
MΩ
COMP
STORE
LOCK
61-797
>600V
COM
OFF
50V
V
100V
250V
500V
1000V
FUSED
CAT. IV
600V
0
LPF
Leak
age
Cu
rre
nt
TEST
TEST
Power-up options:
Press button while turning meter on from OFF position.
Lock button:
display of the software version.
Store button:
reset all stored data.
Blue button:
disable auto power off.
Battery and Fuse Replacement
Refer to the following figure to replace fuse and batteries.
Testing the Fuse
CAUTION:
•
Use only a fuse with the
amperage, interrupt, voltage,
and speed rating specified.
• Fuse rating : Fast, 315mA,
1000V, Min Interrupt
Rating 10000A.
• Replace the batteries as
soon as the low battery
indicator “ ” appears,
to avoid false readings.
• When the battery is too low
for reliable operation, the meter
displays “bAtt.” The meter will
not operate at all until the battery
is replaced.
• 1.5V x 4 alkaline batteries.
• Remove the test leads before
testing the fuse.
• If the display is reading FUSE,
the fuse is bad and should be
replaced.