System set up
Example of system setup when TEK-SCOPE Software is used:
DO NOT CONNECT THE DEVICE BOTH TO A MONITOR AND PC TOGETHER.
If the computer and the monitor are used together, like extended or dual, the algorithm for magnification/
measuring will be effected.
With the following screens sizes: 20”(1600x900); 22”(1920x1080); 22”(1680x1050); 24”(1920x1080);
24”(1920x1200) measuring will have a tolerance of 1-2% without calibration.
Measuring
Put your object under the camera.
Choose the magnification by the software buttons.
Notice: using the remote control, the Zoom factor will be not updated and measuring will be wrong.
DO NOT USE REMOT CONTROL WHEN CONNECTED TO THE PC BUT USE THE SOFTWARE TOOLBAR.
Press the draw button
. Press LB (Left button) on your mouse, select the horizontal
or vertical
button
and click the starting point. Then go to the second measuring point and press LB on your mouse. The measure
value will appear on the screen.
Calibration
If you need a more accurate measuring, the calibration should be done.
Put a ruler horizontally on the subject you wish to measure. Verify that your magnification is the one you want,
select the drawing mode
, right click on the mouse and select “calibration”. A message box will appear, telling
if you can use 1 mm or 10 mm calibration. The calibration setting (1mm or 10mm) depends on the magnification
you use. Press the LB on your mouse to select the length of 1mm or 10 mm. A yellow line will confirm the length