User Manual
34
7.2.8
Basic script example
The following commands can be a minimal script to perform a basic set up of the unit
and read the data
Command
Comment
*IDN?;
WAV 1550;
pow 14;
stat 1;
busy?;
Query idn string of unit
Sets Laser 1-1-1 Wavelength to 1550nm
Sets output Power of laser 1-1-1 to 14 dBm
Enables Laser 1-1-1 output
Queries if the laser port has finished tuning and has settled on target
settings that were set before
Screenshot of script example result using Putty as a terminal Program:
<ENTER> was used to execute the command. Alternatively, use “;”.
<ENTER> was used to
execute the command. Alternatively, use “;”.
If both is used, it
is interpreted as 2 commands of which the second one is empty so that the first
command is executed but the second produces an error since empty.
7.3
Laser port addressing scheme
Laser ports are addressed by a three-level port identifier that allows easy
identification of the port and are issued as parameters with according commands.
Each level parameter is separated by a “,” character.
If not laser port address is added to the command, Laser port 1,1,1 is addressed
automatically, i.e.
SOUR:WAV?
Queries the wavelength setting of laser port 1.
Identifier
Description