The Length provides the distance between the two selected points. In this case, 3.032 um.
Since the known pitch is 3.00 um, the correction factor is 3.00/3.032 or 0.99. In this case, the
pitch is within 1% of the known value.
With the calibration factor calculated, the Dimensions and Units tool is found under
Data
Process > Basic Operations > Dimensions and Units.
In the ‘X Calibration Factor’ input area,
enter the calibration factor. These steps are outlined in Section 9.1.3.1 Apply calibration factor
using Dimensions and Units Tool.