background image

 
 
© ICSPI Corp. 2007-2021 

nGauge AFM User Manual 2.0 

72 

 

 

Where to select plateau and valley heights

 

 

9.1.3.

 

Single-Point Vertical Calibration Example 

To calculate the calibration factor for single-point calibration, take the actual height of the 
calibration and divide it by the measured height. 

For example, if your calibration grating possesses 100 nm-tall features and the measured line 
profile shows 200 nm (0.2 um), then your vertical calibration factor is 100/200, or 0.5. (This is an 
extreme example.) 

From the image above, it is shown that the average difference between peak and valley is 140.3 
nm. The grating size is 110 nm, so the calibration factor is then 0.78 nm.  

9.1.3.1.

 

Apply calibration factor using Dimensions and Units Tool 

With the calibration factor calculated, the Dimensions and Units tool is found under 

Data 

Process > Basic Operations > Dimensions and Units. 

In the ‘Z Calibration Factor’ input area, 

enter the calibration factor calculated earlier. 

Содержание nGauge

Страница 1: ...nGauge Atomic Force Microscope User Manual Version 2 0 For Software and Firmware versions 1 0 0 0...

Страница 2: ...Operation 13 5 1 Start up 13 5 2 Chip Installation and Sample Placement 14 5 3 Software Overview 17 5 4 Sweep 17 5 5 Approach 19 5 6 Scan 23 5 7 Scanning Recommendations 29 5 8 Improving Performance...

Страница 3: ...ICSPI Corp 2007 2021 nGauge AFM User Manual 2 0 3 8 3 Menu Button Advanced Controls 58 9 Appendix 61 9 1 Gwyddion Supplement 61...

Страница 4: ...ment Use only the power supply provided The driving electronics are designed to accept 7 5 V DC and an internal fuse prevents the input current from exceeding 1 A The stage contains moving parts that...

Страница 5: ...the tip facing down in the storage box Use tweezers to pick them up The AFM chip can be placed on a flat surface with the tip facing up The AFM chip can be picked up by hand as long as you don t touc...

Страница 6: ...the software go to the Approach page and click and hold the Retract button Probe tip is susceptible to damage when engaged Do not touch the sample chip or unit when the tip is approaching or in contac...

Страница 7: ...Conventional AFMs typically use a probe with a sharp tip on a silicon cantilever A laser beam is reflected off the backside of the cantilever onto a photodetector that measures any small displacement...

Страница 8: ...m These screws are removable and the stage can be customized 3 3 2 Hole in base plate The base plate features a hole for mounting the nGauge AFM onto an optical table 3 4 nGauge AFM Capabilities 3 4 1...

Страница 9: ...o 10 m in Z peak to valley with 1 nm RMS noise in the vertical direction 3 4 4 Sample Types Any solid sample can be imaged including both conductive and non conductive but the nGauge AFM cannot operat...

Страница 10: ...ow frequency vibrations such as those caused by people walking by doors opening and closing and building vibrations However it is best to set the system up on a solid stable surface away from as many...

Страница 11: ...0 11 4 2 Basic Set up Follow these steps to set up your nGauge AFM system 1 Remove the components from the carrying case Set the stage upright on a flat surface 2 Plug the power supply into a wall ou...

Страница 12: ...cspicorp com resources To install the software run the executable installer nGauge_setup_x x x x_windows exe 4 3 1 Automatic Updates If a network internet connection is available the software will aut...

Страница 13: ...on net download php See Section 6 1 for more details on how to use Gwyddion 5 Basic Operation 5 1 Start up Launch the software by double clicking on the nGauge program on your desktop or searching for...

Страница 14: ...Chip Installation and Sample Placement Once the AFM has been set up and the software installed you are ready to install a chip 1 Completely unscrew and remove the thumb nut from the stage 2 Open the...

Страница 15: ...e sample platform should be retracted sufficiently to avoid any contact between the tip and platform If the sample platform is 10 mm away it is recommended to go to the Approach page in the software a...

Страница 16: ...and holding down the Retract button until the platform is low enough Position the sample so that the area to be scanned is directly under the tip You can also remove the chip before placing the sample...

Страница 17: ...nu are not required for normal operation and only advanced users should use these controls More details are available in Section 8 3 Menu Button Advanced Controls 5 4 Sweep Click the Sweep button to p...

Страница 18: ...n the Sweep button is clicked the resistances of the MEMS scanners of the AFM chip are checked There are five scanners on the AFM chip X1 X2 Y1 Y2 and Z A high or low resistance value indicates that t...

Страница 19: ...for the Motor status and the Approach status Ensure that the target area of your sample is below the nGauge AFM tip Click the Approach button to initiate the automatic approach of the sample to the pr...

Страница 20: ...e tip is engaged with the sample and the tip is not on the target area you must click on Disengage and then click and hold on Retract to move the sample platform away from the tip before re positionin...

Страница 21: ...5 2 False Engage for more information If the amplitude does not track the setpoint it is noisy or far away from the setpoint refer to Section 7 3 6 Setpoint Not Tracked 5 5 2 False Engage If the ampli...

Страница 22: ...ne Z and Vertical Scan Range 5 5 4 Fine Z Drift After approaching you may notice that the Fine Z drifts upwards or downwards Usually this is due to sample relaxation such as polymer relaxation The MEM...

Страница 23: ...4 Scan settings 5 File settings 6 Advanced Controls Click Scan to begin scanning The image will begin to build up line by line The color in the image represents the height of the features a bright go...

Страница 24: ...ed simultaneously topography offset phase and error In addition both the forward and reverse scans are collected for each type 5 6 3 1 Topography scan The topography scan displays data in nanometer nm...

Страница 25: ...ossible 5 6 4 Scan Area The scan area is where the scan will be displayed The top of the scan area shows the size of the scan in micrometers or nanometers The legend to the right of the scan area chan...

Страница 26: ...it will take longer to complete Default is 256 lines 8 16 32 64 128 256 512 1024 Scan range Size of the scan Default is 10 m 1 20 m for area 1 60 m for line Time spent at each point The amount of time...

Страница 27: ...your computer s configuration Windows can occasionally store the data in the AppData folder C Users username AppData Local VirtualStore Program Files x86 ICSPI nGauge All 6 images topography phase er...

Страница 28: ...ied Leveling Direction switches between adding or subtracting a tilt and is device specific Different types of AFM chips may require different Leveling Directions Forward Forward Reverse Leveling If t...

Страница 29: ...ine Z Warning Disabling the controller while engaged may damaged the tip N A N A 5 7 Scanning Recommendations 5 7 1 Changing the scan size Changing the scan size considerably for example from 20 micro...

Страница 30: ...of its range because the tip goes back to its initial lateral position The end of the scanner range is 0 V or 3 3 V on the Fine Z page If the tip reaches either of these extremes data can no longer be...

Страница 31: ...ample to ensure that the tip comes into contact with the sample 2 The tip and cantilever assembly bends away from the chip at a certain angle see image of nGauge AFM chip in Section 3 2 MEMS AFM or AF...

Страница 32: ...e Leveling setting is set correctly These scans show the difference in the tilt when the leveling direction and leveling magnitude are changed Each scan is opened in Gwyddion and a line profile is tak...

Страница 33: ...er words lowering the Z Fine PGA makes the tip more prone to damage as the Z scanner can more easily go out of range Before the Z Fine PGA is decreased it is recommended that the Leveling is set to an...

Страница 34: ...setpoint is recommended when scanning materials that may move around such as particles that are not strongly affixed to the surface substrate A high setpoint may be 75 of the drive amplitude For examp...

Страница 35: ...as flat an error image as possible 5 8 3 1 PI Settings Example As an example three different PI settings are used during the scan of a calibration grating The line profiles of the topography are show...

Страница 36: ...s should be as flat as possible 5 9 Advanced Scan Features There are several advanced scan features that are explained in this section 5 9 1 Line Profile View A profile view can be found below the ima...

Страница 37: ...into an area of interest To zoom into a region of interest you can click and drag across the displayed image to select an area to zoom in Click the Scan button again after the selection has been made...

Страница 38: ...r width of a specific feature or the RMS roughness of the surface These measurements are best acquired using Gwyddion a powerful free and open source AFM post processing program Gwyddion can be downlo...

Страница 39: ...ct Data Align Rows If each horizontal line in the image does not quite line up with the ones above and below it this function will subtract the height differences line by line Median of differences wo...

Страница 40: ...ches for X Y calibration are required Each chip will have slightly different calibration factors An example on how to process scans collect line profiles and apply calibration factors is available in...

Страница 41: ...500 nm tall it is important to have a 100 nm and a 500 nm calibration grating that you can use to calibrate 6 2 2 Lateral Calibration For XY calibration it is important to have calibration gratings wi...

Страница 42: ...Waiting for serial connection will be displayed instead of AFM Ready and the Firmware version will be x x x x Follow these steps to resolve the issue 1 Unplug and then plug in the USB and or power Re...

Страница 43: ...lose the nGauge AFM software and navigate to C Program Files x86 ICSPI nGauge firmware ISP If the firmware folder does not exist download it from the ICSPI website https www icspicorp com downloads Do...

Страница 44: ...M unit is plugged in to the computer and plugged in to power but is not recognized in Device Manager by your computer contact ICSPI for support 7 2 Sweep Troubleshooting 7 2 1 Broken device or device...

Страница 45: ...n 7 2 3 Inspecting AFM Chips 7 2 1 2 Disabling the resistance check Occasionally the resistance check routine is overly sensitive for the Z scanner Follow these steps 1 Navigate to the advanced menu b...

Страница 46: ...an error proceed to 5a If the resistances are above 250 ohm after following these steps one or more of the lateral scanners of your device may have been damaged 5a If the frequency sweep is smooth not...

Страница 47: ...o 55 and is too high Decrease the value and click on Sweep Continue to change the Cantilever Drive Amplitude and re Sweep until you are satisfied 7 2 3 Inspecting AFM Chips If the frequency sweep cont...

Страница 48: ...notification on the splash screen The sample platform can only be lowered so far before a limit switch prevents further travel Once the limit switch is triggered you will not be able to retract any m...

Страница 49: ...s flat and noisy tighten the thumb nut on the AFM chip If the problem persists the AFM chip may be damaged If the maximum value of the peak is less than 2V click on Show Advanced Controls and increase...

Страница 50: ...that the maximum of the amplitude peak is as high as possible depending on the type of chip The recommended maximum cantilever drive amplitude is 3 V for aluminum tips and 2 V for diamond tips Decreas...

Страница 51: ...ne Z Drift and Fine Z Not Centered It is possible that the sample topography is not centered around the middle of the Fine Z range which could result in image clipping The Approach tab displays a grap...

Страница 52: ...otor Speed by dragging the indicator to the right 7 4 Scan Troubleshooting 7 4 1 Drift If there are large line to line variations in height the sample may not be secure For instance if imaging a light...

Страница 53: ...essing in Gwyddion using the Mark scars or Limit Range tools It is usually caused by electrostatic effects from samples that are electrically resistive 7 4 2 3 Wave pattern If the topography of the sa...

Страница 54: ...shows that it reached 3 3 V near the end of the scan There are two main settings to change to ensure that the Fine Z does not reach either end of the scan range 1 Leveling refer to Section 5 7 2 1 Lev...

Страница 55: ...ient for normal imaging conditions However each tab in the navigation contains advanced controls that can extend the functionality or control of the scan In addition there is a side menu available by...

Страница 56: ...rease the tip oscillation amplitude Default value is 35 Different chips may require different cantilever drive amplitudes 8 1 3 Manual Sweep Controls The manual sweep controls are not required for nor...

Страница 57: ...d the coarse Z positioning Parameter Description Proportional The proportional gain coefficient of the PI controller See Section 5 8 3 Proportional and Integral Control for details on tuning the contr...

Страница 58: ...the platform move up and Retract makes the platform go down clicking either Approach or Retract takes a single step while clicking and holding Approach or Retract moves the platform continuously PID C...

Страница 59: ...frequency To set a new frequency change both start and end frequencies to the desired frequency Safety Configurations Encapsulates various safety features Safety thresholds and features may be modifi...

Страница 60: ...rmware update Firmware update files are binary bin files and can be downloaded from our support page Enter Bootloader erase the firmware program to allow for firmware updates or flashing custom firmwa...

Страница 61: ...is not raw In order to do perform calibration scans must be saved as gsf in nGauge software and opened in Gwyddion 9 1 1 1 Change the Color Changing the color is optional but the default for AFM is g...

Страница 62: ...tilt in Y by changing the Levelling setting This will physically change the angle at which the tip scans across the sample To simply correct for the tilt in a scan you can also click on the Level Imag...

Страница 63: ...n under Tools To use the plane fit tool the user clicks on three points that should have the same height and Gwyddion fits a plane through these three points We re going to place these three points in...

Страница 64: ...ICSPI Corp 2007 2021 nGauge AFM User Manual 2 0 64 Three Point Level tool Three points selected in the troughs of the grating...

Страница 65: ...to measure the step heights 9 1 1 5 Line Profile Click on the Line Profile tool When the Line Profile tool is pressed a blank graph will show up In order to show a line profile click onto the scan ima...

Страница 66: ...easing this will increase the area over which the profile of the grating is averaged Increasing this value will result in a smoother profile as more measurements are used to determine the profile When...

Страница 67: ...new graph select Graph Critical Dimension This allows for robust step height measurements Once the tool is open select Step Height Positive Negative from a dropdown on the right hand side Negative is...

Страница 68: ...the appropriate step height calculator is selected click and drag a range on the image which matches the guide picture Once this is selected hit Fit The parameters will fill out with values Looking b...

Страница 69: ...data obtained with the critical dimension tool On the line profile graph select the Measure Distances in Graphs button in the bottom right corner A window will pop up This will allow you to gather dat...

Страница 70: ...th likely two peaks This is a density function of the heights One peak will be found at the height of the plateaus of the calibration grating and one peak will be found at the height of the valleys If...

Страница 71: ...ICSPI Corp 2007 2021 nGauge AFM User Manual 2 0 71 1D Statistical Functions button Measure distances button...

Страница 72: ...ed line profile shows 200 nm 0 2 um then your vertical calibration factor is 100 200 or 0 5 This is an extreme example From the image above it is shown that the average difference between peak and val...

Страница 73: ...ready existing image To see the results firsthand create a line profile and make sure that the heights of the grating are as desired 9 1 4 Multi Point Vertical Calibration If there is a scan for which...

Страница 74: ...ation data at 20 nm and at 110 nm and use these measurements to correct your 50 nm tall feature heights 9 1 4 2 Create a Linear or Polynomial Correction Function Follow the same post processing steps...

Страница 75: ...on More complicated functions such as a polynomial function can be applied to a datafield the Gwyddion term for a scan by using Data Process Multidata Arithmetic The function calculated can be applied...

Страница 76: ...ation can be performed in the X and Y directions A grating with a known pitch is required Often the same grating can be used for vertical and lateral calibration For example to calibrate in the X dire...

Страница 77: ...ICSPI Corp 2007 2021 nGauge AFM User Manual 2 0 77 Use the Measure distances in graph tool in the bottom left Then select a point at the beginning of each of the line structures...

Страница 78: ...factor is 3 00 3 032 or 0 99 In this case the pitch is within 1 of the known value With the calibration factor calculated the Dimensions and Units tool is found under Data Process Basic Operations Dim...

Страница 79: ...User Manual 2 0 79 ICSPI Corp Integrated Circuit Scanning Probe Instruments Corp The AFM on a chip Company 248 Corrie Cres Waterloo Ontario N2L 6E1 Canada Telephone 1 289 236 0204 Email info icspicorp...

Отзывы: