8
FOB-BA-e-2124
KERN FOB-LM:
KERN
FOB 1K-4LM
FOB 3K-3LM
Item no./ Type
TFOB 1K-4LM-A
TFOB 3K-3LM-A
Readability (d)
0.5 g
1 g
Weighing range (max)
1.5 kg
3 kg
Verification value (e)
0.5 g
1 g
Verification class
III
III
Minimum weight (min)
10 g
20 g
Reproducibility
0.5 g
1 g
Linearity
± 1.5 g
± 2 g
Recommended adjustment
weight, not added (class)
1 kg (M1)
2 kg (M1)
Range for zero point adjustment
(A/D converter value)
3000 - 120000
3000 - 120000
Range for adjustment with
calibration weight (A/D converter
value)
30000 – 50000
40000 – 65000
Warm-up time
10 min
10 min
Weighing unit
g
Stabilization time (typical)
2 sec.
Operating temperature
+ 5° C …. + 35° C
Humidity of air
25% - 95% (non-condensing)
Housing (B x D x H) mm
285 x 255 x 90
Weighing plate, stainless steel
(mm)
252 x 200 x 14
Weight kg (net)
3.8 kg
Auto off
2 min
Battery
4 x 1.5 V AA
Service life background illumination ON: 48 h
Service life background illumination OFF: 66 h
Input voltage Appliance
12 V / 500 mA
Input voltage Mains adapter
230 V / 50 Hz
IP protection
IP67