FOB-BA-e-2124
41
10.5.2 Automatic switch-off function „AUTO OFF“
in stand-by mode
The balance offers the possibility of the automatic switch-off, when the balance is in
stand-by mode. The switch-off time may be selected as follows:
Models FOB-S, FOB-NL:
Balance display
Operation
ò
FOB-S:
•
In weighing mode, press and hold the
TARE
button
for approx. 3 seconds.
„SEt“, followed by „Unit“ is displayed
FOB-NL:
•
Switch off the balance
•
Press and hold the
ON/OFF
key and
simultaneously press the
TARE
key 3 times and
then release both keys.
•
Press the
UNIT
key, "AOFF" is displayed
ò
ò
ò
ò
•
Press the
TARE
-key, current setting appears.
•
Press the
UNIT
key to select the desired setting:
120s = Auto off after 120 seconds
180s = Auto off after 180 seconds
240s = Auto off after 240 seconds
300s = Auto off after 300 seconds
OFF = Auto off switched off
•
Press the
TARE
key to confirm your selection,
"AOFF" is displayed
•
Press the
UNIT
key, "CAP" is displayed