30
FOB-BA-e-2124
9.4 Hold function (FOB, FOB-NL/-LM/ models only))
The balance has an integrated standstill function (mean value calculation).
With this it is possible to weigh unstable samples to be weighed exactly.
There are 2 possibilities:
Balance display
Operation
•
Turn on balance by pressing the
ON/OFF
key.
The balance will carry out a self-test.
Wait for
„0“
display
(example)
•
Put the weighed item in place and press the
HOLD
key; "HOLD" will appear and a small triangle
flashes. During this time, an average value is
recorded.
•
The triangle will stop flashing, the stability display
appears and the weight value obtained is displayed.
•
By pressing the
HOLD
button several times, the
balance returns to weighing mode. The triangle
disappears.
9.5 Display background illumination (FOB, FOB-NL/-LM/ models only)
The balance has the ability to turn the backlight of the display on or off.
Press the
TARE
key for approx. 3 sec.