Firmware Version 1.2
MultiSystem 4070
72
Operating software
DEU
Using the digital signal output, you can carry out event-dependent external
control.
You are able to define up to 5 parameters here.
Measurand
shows the internal measurand of the output
Channel name
You can assign an arbitrary name here
Mode
Source of the triggering event;
•
INACTIVE
Trigger off
•
CHANNEL
Channel is monitored for the occurrence of the triggering event,
•
SP-TRIG
Trigger is set if trigger was detected during saving.
This allows multiple measuring instruments to be synchronized:
– Master: Saving of triggering event X (e.g. p1>200) – trigger output:
SP_TRIG;
– Slaves: Saving of triggering event E1
•
MANUAL
: the trigger output is switched manually by pressing a key
Reference channel
Select the channel to be used as reference channel.
Condition
for trigger input
OFF
/
ON
for measurement channels
GREATER THAN
/
LESS THAN
Value
for measurement channels, e.g. 200