Firmware Version 1.2
MultiSystem 4070
103
Operating software
DEU
9
Press
to toggle between:
•
–
The settings in the menu can be changed.
•
LOCKED
The settings in the menu are locked and cannot be changed. When trying
to apply a change in a locked menu, a corresponding warning will be dis-
played after pressing
.
10
Confirm changes and exit dialog:
Calibration setting
The calibration interval is the period of time after which the measuring instru-
ment is to be re-calibrated by the manufacturer.
You can define the calibration interval.
The measuring instrument is also ready for use if no calibration interval is set.
Hinweis
If a release code is set in the
Security
dialog, you must enter this release code
before attempting to release menus in the dialog or lock additional menus.
Resetting the instrument deletes the password in the
Security
dialog and
releases all locks set on the menus.
OK
Confirms input/saves change.