HRT-710 User Manual (Version 1.24, June/2015) PAGE: 31
Word 4
:
0x1B (Byte 9) 0x10 (Byte 8)
Word 5
:
0x1B (Byte 11) 0x00 (Byte 10)
Word 6
:
0xE8 (Byte 13) 0x97 (Byte 12)
The HART command 0 format is 2 bytes response code and 12
bytes data.
[ Response code1 ]
Byte 0
:
0x00
means ―No Error‖
[ Response code2 ]
Byte 1
:
0x10
means ―More Status Available‖
[ Response data bytes of command 0 ]
Byte 2
:
0xFE
Constant value
Byte 3
:
0x3F
Manufacturer ID, 0x3F = ―Eckardt‖
Byte 4
:
0x04
Manufacturer’s device ID
Byte 5
:
0x08
Number of preambles needed in the request
Byte 6
:
0x05
Command set revision number
Byte 7
:
0x01
Transmitter specific revision code
Byte 8
:
0x10
Software revision
Byte 9
:
0x1B
Hardware revision
Byte 10
:
0x00
Flags
Byte 11~13
:
0x1B 0x97 0xE8
Device ID number