background image

1. De-pressurize and Vent System Prior to Installation or Removal.
2. Confirm Chemical Compatibility Before Use.
3. DO NOT exceed Maximum Temperature or Pressure Specifications.
4. ALWAYS Wear Safety Goggles or Face-shield During Installation and/or Service.
5. DO NOT Alter Product Construction.

If  leaking  is  observed  from  the  retaining  cap,  it  indicates 
defective  or  worn  o-rings  on  the  sensor.  Do  not  attempt  to 
correct by further tightening.

WARNING!

Indicates a potential hazard. Failure to follow all 
warnings may lead to equipment damage, injury, or 
death

Warning | Caution | Danger

Overtightening may permanently damage product 
threads and lead to failure of the retaining nut.

Hand Tighten Only

Failure to follow these instructions may result in
the sensor being ejected from the pipe!

Use of tool(s) may damage product beyond repair 
and potentially void product warranty.

Do Not Use Tools

Highlights additional information or detailed 
procedure.

Note | Technical Notes

Safety Information

Do Not Remove Under Pressure

Please ensure that the Instruments are not to be subject to water hammer or pressure spikes!
Always Pressure Test System with H

2

O Prior to Initial Start-Up

Before  Before  installation  be  certain  the  appropriate  instrument  has  been  selected  considering  operating  pressure,  full  scale 
pressure,  wetted  material  requirements,  media  compatibility,  operating  temperature,  vibration,  pulsation,  desired  accuracy  and 
any other instrument component related to the service application including the potential need for protective attachments and/or 
special installation requirements. Failure to do so could result in equipment damage, failure and/or personal injury. Ensure only 
qualified personnel personnel are permitted to install and maintain this instrument.

Sensor may be under pressure, take caution to vent system prior to installation or removal. 
Failure to do so may result in equipment damage and/or serious injury

Pressurize System Warning

TK Series can be installed in a horizontal or vertical direction. Please ensure enough length of straight pipe to avoid turbulence 
that can effect readings.

A Bag Filter or Y Strainer Filtering Device upstream to Avoid the Paddle Wheel from being damaged by the solids or fibers
- max 10% Particle Size - Not to Exceed .5mm Cross Section or Length.

Please do not flush the pipe after the Flow Meter is installed with Compressed Air this may damage the ceramic shaft and will 
Void Warranty

Please Ensure Full Pipe

Min 10x Pipe Diameters Upstream 3x Pipe Diameters Downstream.

Always utilize the most appropriate PPE during installation and service of Truflo products.

Personal Protective Equipmet (PPE)

www.iconprocon.com | [email protected] | 905.469.9283

03

In-Line Paddle Wheel Flow Meter

TKP 

|

 TKM SERIES 

(Quick Start)

Содержание truflo TKP Series

Страница 1: ...Paddle Wheel Flow Meter TKP TKM SERIES Quick Start Operating Manual www iconprocon com sales iconprocon com 905 469 9283 In Line Paddle Wheel Flow Meter TKP TKM SERIES Quick Start 360...

Страница 2: ...ettings K Factors for TK Flow Rates Pressure vs Temperature Psi H2O Non Shock Procedure to Rotate Display Dimensions Installation Positions Warranty Returns and Limitations 03 04 05 05 06 06 07 07 08...

Страница 3: ...ure full scale pressure wetted material requirements media compatibility operating temperature vibration pulsation desired accuracy and any other instrument component related to the service applicatio...

Страница 4: ...m s nominal Supply Voltage 5 to 24 VDC 10 regulated Supply Current 1 5 mA 3 3 to 6 VDC 15 Hz per ft s nominal 3 VDC 20 mA 6 to 24 VDC Wetted Materials Sensor Body PVC Dark PP Pigmented PVDF Natural O...

Страница 5: ...Flow Totalizer NPN Pulse Diagram Yellow Grey with RS485 Only Black Wire can be Changed for Flow Total Limit Output or Unit Volume Pulse Output Brown Black White Yellow Grey Blue TKP Flow Meter Output...

Страница 6: ...STEPS Home Screen Press Key 3 sec Step 1 Programming of Lock Out Feature Press Key Step 2 Programming Units of Flow Press Key Step 3 Coefficient of Flow Volume Press Key Step 4 No Programming Required...

Страница 7: ...Output Flow Rate Pulse Press Key 3 sec OP1 Step 4 Flow Rate Program Time Delay for NPN Pulse OP1 on Initial Start Up Range 0 9999 Secs Time Delay of Pulse Output Flow Rate T2 20 Secs Default Flow Rate...

Страница 8: ...e Size O D LPM GPM LPM GPM 0 3m s min 10m s max DN15 DN20 3 4 DN25 1 DN40 1 DN50 2 2 DN80 3 DN100 4 3 5 1 0 5 0 1 5 9 0 2 5 25 0 6 5 40 0 10 5 60 0 16 90 0 24 125 0 33 120 32 170 45 300 79 850 225 135...

Страница 9: ...Degrees Lower Display Tighten the Allen screws Snug Tight Do Not Over Tighten Pipe Size d inch D inch C inch L inch H inch 4 09 0 05 4 17 0 05 4 30 0 05 5 02 0 05 5 56 0 05 5 48 0 05 6 12 0 05 6 76 0...

Страница 10: ...tream to Avoid the Paddle Wheel from being damaged by the solids or fibers max 10 Particle Size Not to Exceed 5mm Cross Section or Length Please do not flush the pipe after the Flow Meter is installed...

Страница 11: ...ical mechanical or chemical damage due to improper accidental or negligent use 3 have been modified or altered 4 anyone other than service personnel authorized by Icon Process Controls Ltd have attemp...

Страница 12: ......

Отзывы: