FMX 400 P SERIES IR CAMERA USER MANUAL
15
Site:
_______________________________
System:
___________________________
Contact Person:
__________________
Date:
______________________________
**All those tasked with the initial setup of the temperature screening equipment must be
present during this training.
***This portion of the validation requires a live video feed with the instructors.
10-1 The Screening Area
□
Ensure that this area is free from:
•
Any direct or indirect (reflected) sunlight
•
Warm or cold conductive airflow
◦
HVAC vents/intakes
•
Any radiant energy from electrical sources
•
Direct or indirect lighting on individual being screened
□ Room temperature is 20 °C - 24 °C (68 °F - 75 °F)
□ Relative humidity is within 10% – 50%
10-2 The Screening Background
□
Placement is parallel and perpendicular to screening camera
•
At a distance of ~12 feet
□
Consists of a non-IR reflective homogeneous background
•
Minimum size is 10’ x 8’
•
White to gray flat finish
□
The Temperature Reference Source is in front of background (direct line of sight to the
screening camera) and is framed within the screening image
10.
Site and System Setup Validation Checklist
Continues