TURNING THE TONOMETER ON AND LOADING THE PROBE

Pressing the Select button

Pressing the Measure button

1.

 Press the Select or Measure button to turn the tonometer on. Illustrations of these alternative ways of starting:

     Remember to use the wrist strap

2.

 

Open the probe tube by removing the cap and  

  insert the probe into the probe base as shown in  

  the image.

3.

 When the Play symbol appears on the display,  

  the tonometer will be ready for measurements

2.

3.

To change the setting for different species 

press the 

Select button to access the menu. Press Navigation 

buttons and select SETTING by pressing the Select 

button. Toggle the settings by Navigation buttons and 

select a specie setting by pressing Select button. To 

measure return to MEASURE by navigation buttons.

MEASUREMENT

1.

 Bring the tonometer near the patient’s eye. (It’s not necessary to look at the display during the measurement 

  since the device gives sound signals about errors etc.)
Avoid excessive restraining, as it may alter the IOP. Patient’s head should be held as lightly as possible; 

be careful not to put pressure on the neck or the eye ball. If a collar is worn; make sure it is not too tight 

or remove it for the measurement.