PARTS OF THE TONOMETER
• Fill the probe base cleaning container or other clean container
with 70-100 % isopropyl alcohol.
• Turn the power off.
• Unscrew the probe base collar and remove the probe base by
tilting the tonometer downwards.
• Insert the probe base into the probe base cleaning container
or other clean container filled with 70-100% isopropyl alcohol.
• Let soak for 5-30 minutes.
• Remove the probe base from alcohol.
• Dry the probe base by blowing clean canned or compressed air
into the hole in the probe base. This will additionally remove
possible residual dirt.
• Insert the probe base into the tonometer.
• Screw the collar in, to lock the probe base.
Please refer to the instruction manual for more detailed instructions
of changing and cleaning the probe base.
CHANGE/CLEAN THE PROBE BASE REGULARLY TO AVOID
UNNECESSARY SERVICE.
www.tonovet.com
Measuring support
Probe base and probe base light
Collar
Navigation buttons
Measuring support adjusting wheel
Display
Measure button
Select button
Please refer to the instruction manual
for additional information. (USB)
CLEANING THE PROBE BASE
Replace the probe base every 12 months. Clean the probe base every 6 months. Clean or replace the probe
base if the error message Clean Change is displayed. Make sure that the probe base is completely dry
before using it.