PARTS OF THE TONOMETER

• Fill the probe base cleaning container or other clean container  
  with 70-100 % isopropyl alcohol. 
• Turn the power off. 
• Unscrew the probe base collar and remove the probe base by  
  tilting the tonometer downwards.
• Insert the probe base into the probe base cleaning container 
  or other clean container filled with 70-100% isopropyl alcohol. 
• Let soak for 5-30 minutes.
• Remove the probe base from alcohol.
• Dry the probe base by blowing clean canned or compressed air  
  into the hole in the probe base. This will additionally remove 
  possible residual dirt. 
• Insert the probe base into the tonometer. 
• Screw the collar in, to lock the probe base.

Please refer to the instruction manual for more detailed instructions 
of changing and cleaning the probe base.

CHANGE/CLEAN THE PROBE BASE REGULARLY TO AVOID 
UNNECESSARY SERVICE.

www.tonovet.com

Measuring support   

Probe base and probe base light   

Collar 

Navigation buttons   

 

 Measuring support adjusting wheel
 Display
 Measure button
 Select button

   

Please refer to the instruction manual 

for additional information. (USB)

CLEANING THE PROBE BASE

Replace the probe base every 12 months. Clean the probe base every 6 months. Clean or replace the probe 
base if the error message Clean Change is displayed. Make sure that the probe base is completely dry 
before using it.