Green probe base
light indication
ONE SHORT BEEP
Indicates a correct single measurement.
DOUBLE BEEP
Indicates an incorrect single measurement. (See error messages from full manual).
LONG BEEP
Indicates that six succesful measurements are done and IOP result is available at the display.
Icare
TONOVET
Plus tonometer Quick Guide
ENGLISH
2.
Keep the
probe horizontal and pointing perpendicularly to the center of the cornea
. The distance
from the tip of the probe to the patient´s cornea should be
4 - 8 mm (5/32 - 5/16 inch)
. The probe
base light must be green (indicating correct measuring position).
3.
Press the Measure button to measure IOP.
Probe will contact the eye during
measurements: Press and hold the button to collect 6 measurements in a row
OR press the button briefly 6 times to collect 1 measurement at the time.
DISPLAY DURING MEASUREMENTS
Before
After the second
measurement
After the sixth
measurement
After the sixth
measurement but with
too much variation**
After the sixth
measurement but with
acceptable variation*
*
Measurement should be repeated if the result is over 25 mmHg
**
Measurement should be repeated (
NOTE
In very high IOPs, the fluctuation may be higher than normal)