Страница 2: ...ia www labbox com Please provide the customer care representative with the following information Serial number Description of problem Your contact information Warranty This instrument is warranted to...

Страница 3: ...e operation and a favorable application effect Main Technical Parameters Model TWBA 001 001 TWBA 002 001 TWBA 004 001 TWBA 006 001 Voltage 200 240 V 100 120 V Power W 400 500 1000 1500 Temperature Mot...

Страница 4: ...non set state the window displays the prompt SP and the temperature set value Upon re pressing the SET button the window displays the prompt ST and the time set value Press the SET button again and t...

Страница 5: ...the SHIFT AT button for 6 seconds The controller will enter the pre auto tuning state displaying the prompt AT in the upper window and OFF in the lower window Users can use the DEC or INC button to ch...

Страница 6: ...C 6 0 I Integration Time Adjustment of integration function 1 2000 200 d Differential Time Adjustment of differential function 0 1000 200 T Control Cycle The temperature control cycle 1 60 5 Pb Zero P...

Страница 7: ...er enters the constant temperature state 0 1 100 0 C 0 5 SPT Constant Temperature Buzzer Time If in the constant temperature state the buzzer will beep for SPT seconds Note If SPT 9999 the buzzer will...

Страница 8: ...er supply 4 The upper row of the instrument displays the test temperature and the setting temperature is shown in the lower row 5 Briefly press the setting key to enter the setting state Use the SHIFT...

Страница 9: ...ature controller is broken 2 The sensor is broken 3 The set temperature is lower than the water temperature 4 The heating pipe is burnt 1 Replace the instrument 2 Replace the sensor 3 Reset the temper...