background image

 

7

 

Parameter Table 2 

 

Name 

Function Instructions 

(Setting Range) 
Factory Set Value 

Lc  

Password Key 

When Lc=9, enter the next parameters.

 

ndA 

Temperature 
Alarm Mode 

0: Over-temperature alarm only. 
1: Over-temperature alarm and under-temperature 
alarms simultaneously.  

(0

1) 0 

ndc 

Control Mode 

0: PID control. 
1: ON/OFF control. 

(0

1) 0 

dE1 

Upper 
Deviation 

Valid only for ON/OFF control. 

(0

100.0

 ºC

) 0 

dE2 

Lower 
Deviation 

(0

100.0

 ºC

) 0 

ndT 

Timer Mode 

0: No timer function. 
1: Constant temperature timing. 
2: Run timing. 

(0

2)  1 

Hn 

Timer Unit 

0: Minute.   
1: Hour. 

(0

1) 0 

SPd 

Constant 
Temperature 
Deviation 

When 

SP>=(SV-SPd)

, the controller enters the 

constant-temperature state. 

(0.1

100.0

 ºC

0.5 

SPT 

Constant 
Temperature 
Buzzer Time 

If in the constant-temperature state, the buzzer will beep 
for SPT seconds.  
Note: If 

SPT=9999

, the buzzer will beep continuously. 

(0

9999S)  

EST 

Timing Over 
Buzzer Time 

If the timing work is over, the buzzer will beep for EST 
seconds. 
Note: If 

EST=9999

, the buzzer will beep continuously. 

(0

9999S) 

 60 

EH 

Whether to 
continue to 
control after 
timing 

0: Cut off heat-out after timing. 
1: Continue to control after timing. 

(0

1) 0

 

ndo 

 

null 

 

oPn 

 

null 

 

nP 

Maximum 
Power  

Percentage of maximum power heating output. 

(0

100%) 

 100 

Co 

Off Point 

If 

SV>(SP+Co)

, stop the heating output. 

(0

100.0

 ºC

 50.0 

SPL 

Minimum 
Set Point 

The minimum temperature set point. 

(0

50.0

 ºC

) 0

 

SPH 

Maximum Set 
Point 

The maximum temperature set point. 

(SPL

100.0

 ºC

100.0 

 
 
 
 
 
 
 
 
 

Содержание WB01

Страница 1: ...WB01 Thermostatic Water Bath Please read the User Manual carefully before use and follow all operating and safety instructions user manual english...

Страница 2: ...ia www labbox com Please provide the customer care representative with the following information Serial number Description of problem Your contact information Warranty This instrument is warranted to...

Страница 3: ...e operation and a favorable application effect Main Technical Parameters Model TWBA 001 001 TWBA 002 001 TWBA 004 001 TWBA 006 001 Voltage 200 240 V 100 120 V Power W 400 500 1000 1500 Temperature Mot...

Страница 4: ...non set state the window displays the prompt SP and the temperature set value Upon re pressing the SET button the window displays the prompt ST and the time set value Press the SET button again and t...

Страница 5: ...the SHIFT AT button for 6 seconds The controller will enter the pre auto tuning state displaying the prompt AT in the upper window and OFF in the lower window Users can use the DEC or INC button to ch...

Страница 6: ...C 6 0 I Integration Time Adjustment of integration function 1 2000 200 d Differential Time Adjustment of differential function 0 1000 200 T Control Cycle The temperature control cycle 1 60 5 Pb Zero P...

Страница 7: ...er enters the constant temperature state 0 1 100 0 C 0 5 SPT Constant Temperature Buzzer Time If in the constant temperature state the buzzer will beep for SPT seconds Note If SPT 9999 the buzzer will...

Страница 8: ...er supply 4 The upper row of the instrument displays the test temperature and the setting temperature is shown in the lower row 5 Briefly press the setting key to enter the setting state Use the SHIFT...

Страница 9: ...ature controller is broken 2 The sensor is broken 3 The set temperature is lower than the water temperature 4 The heating pipe is burnt 1 Replace the instrument 2 Replace the sensor 3 Reset the temper...

Страница 10: ...10...

Страница 11: ...www labbox com...

Отзывы: