7
Parameter Table 2
Name
Function Instructions
(Setting Range)
Factory Set Value
Lc
Password Key
When Lc=9, enter the next parameters.
0
ndA
Temperature
Alarm Mode
0: Over-temperature alarm only.
1: Over-temperature alarm and under-temperature
alarms simultaneously.
(0
~
1) 0
ndc
Control Mode
0: PID control.
1: ON/OFF control.
(0
~
1) 0
dE1
Upper
Deviation
Valid only for ON/OFF control.
(0
~
100.0
ºC
) 0
dE2
Lower
Deviation
(0
~
100.0
ºC
) 0
ndT
Timer Mode
0: No timer function.
1: Constant temperature timing.
2: Run timing.
(0
~
2) 1
Hn
Timer Unit
0: Minute.
1: Hour.
(0
~
1) 0
SPd
Constant
Temperature
Deviation
When
“
SP>=(SV-SPd)
”
, the controller enters the
constant-temperature state.
(0.1
~
100.0
ºC
)
0.5
SPT
Constant
Temperature
Buzzer Time
If in the constant-temperature state, the buzzer will beep
for SPT seconds.
Note: If
“
SPT=9999
”
, the buzzer will beep continuously.
(0
~
9999S)
0
EST
Timing Over
Buzzer Time
If the timing work is over, the buzzer will beep for EST
seconds.
Note: If
“
EST=9999
”
, the buzzer will beep continuously.
(0
~
9999S)
60
EH
Whether to
continue to
control after
timing
0: Cut off heat-out after timing.
1: Continue to control after timing.
(0
~
1) 0
ndo
null
oPn
null
nP
Maximum
Power
Percentage of maximum power heating output.
(0
~
100%)
100
Co
Off Point
If
“
SV>(SP+Co)
”
, stop the heating output.
(0
~
100.0
ºC
)
50.0
SPL
Minimum
Set Point
The minimum temperature set point.
(0
~
50.0
ºC
) 0
SPH
Maximum Set
Point
The maximum temperature set point.
(SPL
~
100.0
ºC
)
100.0