Parameter Table 1
Name
Function Instructions
(Setting Range)
Factory Set Value
Lc
Password Key
When Lc=3, enter the next parameters.
0
ALH
Over-
temperature
Alarm
If
“
SV>(SP+ALH)
”, the “ALM”
light turns on, the buzzer
sounds, and the heating output turns off.
(0
~
100.0 ºC)
5.0
ALL
Under-
temperature
Alarm
If
“
SV<(SP-ALL)
”
, the
“
ALM
”
light flashes, and the buzzer
sounds.
(0
~
100.0 ºC)
0
P
Proportional
Band
Adjustment of proportional function.
(0.1
~
100.0 ºC)
6.0
I
Integration Time Adjustment of integration function.
(1
~
2000) 200
d
Differential Time Adjustment of differential function.
(0
~
1000) 200
T
Control Cycle
The temperature control cycle.
(1
~
60) 5
Pb
Zero Point
Adjust
When the zero error is relatively larger, an update to
this value may be necessary.
Pb = actual value
–
measure value
(-50.0
~
50.0 ºC)
0
PL
Full Point Adjust
When the full-point error is also relatively larger, an
update to this value may be necessary.
PK = 1000 x (actual value
–
measure value) / measure
value.
(-999
~
999) 0
Addr
null
Loc
Setting Lock
0: Enables setting temperature and time.
1: Disables setting temperature and time.
(0
~
1) 0