Changing bandwidth
Higher bandwidths allow you to accurately measure faster moving targets.
Lower bandwidth reduces the noise on the output and creates higher resolution.
The CPL350 is set to 15 kHz unless specified otherwise in the order and on the calibration sheet received with
the system.
Two switches inside the enclosure set the bandwidth. To change them:
•
Remove the 4 screws that attach the front panel to the cover.
•
Bandwidth is controlled by positions 3 and 4 of DIP-switch S1.
CAUTION:
DO NOT CHANGE ANY OTHER SWITCHES
Set the desired bandwidth according to this table:
S3
S4
Bandwidth
ON
ON
100 Hz
ON
OFF
1 kHz
OFF
ON
10 kHz
OFF
OFF
15 kHz