User Manual |
P5
7
Connecting to the CPL350
7.1
Probe Connections
To Connect the Probes
1.
Insert the probe cable connector into the “Probe” connector on the CPL350.
2.
Rotate until the red dots on the connectors align.
3.
Push the connector in until it clicks.
To Disconnect the Probes
1.
Pull on the knurled barrel of the probe connector which will release the latching mechanism.
CAUTION:
The probe cannot be disconnected by pulling on the cable.
7.2
Analog Output BNC Connector
Single-Ended ±10 V
Analog output is ±10 V. The output voltage becomes more positive as the probe gets closer to the target. The
scaling of the output (i.e. V/mm) is dependent on the probe and calibration specified when ordered. The complete
calibration information is available on the NIST traceable calibration sheet provided with the sensor.
7.3
Power & Differential Analog Output Connector
9-Pin, Male, D-Sub connector with 4-40 mounting hardware.
Pin Assignments
Pin
Connection
1
Ground
2
NC
3
–
15 VDC
4
+15 VDC
5
NC
6
Non-Inverted Analog Output
7
Inverted Analog Output
8
Ground
9
Ground
Power Supply Requirements
The power supply provided with your CPL350 provides the required ±15VDC input power. If using your own
power supply, be aware that high-frequency noise from a switching-type power supply may appear in the output.
For high resolution results, be sure to provide clean power.