Approvals and safety considerations  

The CPL350 is compliant with the following CE standards: 

Safety: EN 61010-1:2010 

EMC: IEC 61326-1:2013, IEC 61326-2-3:2013 

To maintain compliance with these standards, the following operating conditions must be maintained: 

 

All I/O connecting cables must be shielded and less than three meters in length 

 

Use the included CE approved power supply. If an alternative power supply is used, it must have equivalent 
CE certification and provide safety isolation from the mains according to IEC60950 or 61010. 

 

Sensors must not be attached to parts operating at hazardous voltages in excess of 33VRMS or 70VDC 

Use of the equipment in any other manner may impair its safety and EMI protections. 

 

2

 

Overview 

The CPL350 Capacitive Displacement Sensor provides one channel of noncontact displacement measurement 
for precision position and displacement applications. This manual will provide you with all the information you 
require to get the greatest benefit from your system.  

Engineers, technicians, and maintenance personnel will find the CPL350 invaluable for high precision 
measurements of mechanical displacement where contact with the target is not possible nor advisable. Targets 
such as high-speed rotating spindles, or parts that could be damaged or deflected by contacting probes can be 
measured with confidence and accuracy. 

 

3

 

System contents 

Your system includes this manual, the CPL350 Capacitive Sensor electronics, power supply, one probe, and a 
NIST traceable calibration sheet. 

 

4

 

Calibration 

Your system was calibrated with a precision calibrator designed by Lion Precision. The system is calibrated with 
nanometer precision. All of our calibrations are traceable to NIST. 

IBS Precision Engineering recommends that you have your system recalibrated at the factory once a year to 
ensure maximum accuracy. 

 

 

Содержание CPL350

Страница 1: ...www ibspe com User Manual P1 CPL350...

Страница 2: ...TEM CONTENTS 3 4 CALIBRATION 3 5 USING THIS MANUAL 4 6 HELPFUL TECHNICAL SUPPORT DOCUMENTS ONLINE 4 7 CONNECTING TO THE CPL350 5 7 1 PROBE CONNECTIONS 5 7 2 ANALOG OUTPUT BNC CONNECTOR 5 7 3 POWER DIF...

Страница 3: ...or provides one channel of noncontact displacement measurement for precision position and displacement applications This manual will provide you with all the information you require to get the greates...

Страница 4: ...Notes and Application Notes in the Technical Library These documents provide detailed descriptions of the operation and use of the products of IBS Precision Engineering The Technical Library can be ac...

Страница 5: ...mes more positive as the probe gets closer to the target The scaling of the output i e V mm is dependent on the probe and calibration specified when ordered The complete calibration information is ava...

Страница 6: ...puts are each 5 VDC relative to ground the differential output is therefore 10 VDC CAUTION Do NOT connect the outputs to ground or the unit will be damaged Differential Outputs as Single Ended The Non...

Страница 7: ...PL350 is set to 15 kHz unless specified otherwise in the order and on the calibration sheet received with the system Two switches inside the enclosure set the bandwidth To change them Remove the 4 scr...

Страница 8: ...ion Bandwidth 3 dB DIP Switch selectable 100 Hz 1 kHz Operating Temperature 4 50 C Probe Thermal Stability 0 04 F S C Output Impedance 0 Output Voltage Max 13 5 VDC Output Max Current 20 mA Power Outp...

Страница 9: ...0 290 1270 User Manual P9 11 Optional Mounting Flange An optional flange for easy mounting of the CPL350 is available P017 6455 The flange attaches to the bottom of the device with four flathead screw...

Отзывы: