15
When should you carry out a test with the control solution?
•
at least once per week
• if the device has been dropped
•
when you open a new container
• every time when you suspect
of test strips
that the blood sugar measuring
•
if the measured results do not
system is not working correctly
agree with the way you feel
important information about control solutions
-
Only use control solutions included in delivery.
-
Write the date of opening on the bottle.
-
Check the use-by date for the solution. Never use solutions past their use-by date.
-
The control solution should reach room temperature (20°C to 25°C/68°F to 77 °F) before you use it.
-
Shake the bottle of test solution well before you open it. Wipe the first drop away, and then use the
second drop to ensure that you have a good sample for precise measure results.
-
You should use the control solution up within 90 days of opening.
-
Keep the solutions safely closed away at room temperature (below 30°C). Never freeze them!
Warning:
The measured value for the solutions which is printed on the test strip container is only
used to check the measuring system. It is not intended as a recommended value for your
blood sugar level!
checking the system 10
ibp_TD3213_GB 07.01.2005 7:35 Uhr Seite 15