35
Interface
Technical data
Conductivity
0 to 30,0 mS/cm, 4-Pole probe,
Resolution and accuracy:
Range
Resolution
Accuracy
0.0
– 200 µS/cm
0,1 µS/cm
0,6 µS/cm
201 - 1999 µS/cm
1 µS/cm
6µS/cm
2,00
– 30 mS/cm
0,01 mS/cm
0,03 mS/cm
Reference temperature 25° C
Temperature compensation is 2,070 % per Kelvin
Adjustable from 0.00 to 4,00 %/Kelvin
Or selectable from dialysis machine manufacturer preset values.
Temperature (only versions BN, BO, BP and BQ)
Range
0 … 100 °C,
Resolution
0,01 °C
Accuracy
0,1°C
Pressure all versions
Units
mmHg, kPa, mbar, PSI
Versions BO, BP and BQ
Range
-700 … 1900 mmHg, 2.5 Bar
Resolution
0,1 mmHg
Accuracy
0 … 300 mmHg
1 mmHg
Otherwise
2 mmHg
Fluid resistance
yes
Version BM and all other versions with option H)
Range
-700 … 1900 mmHg, 2.5 Bar
Resolution
0.01 mmHg
Accuracy
-299.9 … 300 mmHg
0.5 mmHg
Otherwise
1 mmHg
Fluid resistance
no
HDM97
BH5
Range:
-5 to 5 psi
-0.34 to 0,34 bar
-258 to +258 mmHg
Resolution:
0.01 mmHg
Accuracy:
General 0.05% full scale
Media Compatibility:
Gasses
HDM97
BH15
Range:
- 12 to 15 psi
-0.85 to 1 bar
-650 to +775 mmHg
Resolution:
0.01 mmHg
Accuracy:
General 0.05% full scale
0 to 300 mmHg ± 0.4 mmHg,
Media Compatibility:
Gasses