21
Adjustment and Calibration
Adjustment of customer specific temperature coefficient "USER TC"
Pressing the navigation key
while adjusting the temperature coefficient to "USER TC" the
HDM97
enters a second dialog to adjust the value itself instead of directly going back to measuring
mode.
→
USER TC
→
enters the dialog to adjust the user temperature coefficient set
→
cancel and discard setting
Display:
EDIT TC1
37.00 °C
Temperature associated with the coefficient
2.071 %/K
(Example 2.070, navigation key
has been pressed once)
Keys:
→
increments current value
→
decrements current value
→
Applies the value and leads to the adjustment dialog for the second
pair
→
Discard adjustment and goes on using the last active temperature
coefficient setting
Display:
EDIT TC2
25.00 °C
Temperature associated with the coefficient
2.001 %/K
(Example 2.000, navigation key
has been pressed once)
Keys:
→
increments current value
→
decrements current value
→
Applies the coefficient set. The displays shows "ADJ OK" for 2
seconds
→
Discard adjustment and goes on using the last active temperature
coefficient setting