Gas Incubation System
Table 1 – (continued)
1.5 m (between Column and Heated Lid)
82 g/0.18 lbs
Cable for Humidity Sensor
1.5 m
USB cable
1.8 m
Power supply cable
2.0 m (power supply to wall)
1.2 m (power supply to device)
Gas Connections
Gas tubing material
Polyurethane, hydrolysis resistant
Gas tubing connections
Push-in fittings
Gas Mixer input
ID 4 mm
OD 6 mm
Gas Mixer output and Humidify-
ing Column connection
ID 2.5 mm
OD 4 mm
Connection to the Heated Lid
Male Luer Lock
Gas Input Requirements
Compressed air
Purity class: ISO 8573-1:2010 [1:2:1]
1 bar/14.5 psi optimum (0.8–1.2 bar / 11.6–17.4 psi)
Carbon dioxide (CO
2
)
Purity:
>
99% CO
2
1 bar/14.5 psi optimum (0.8–1.2 bar / 11.6–17.4 psi)
Nitrogen (N
2
)*
Purity:
>
99% N
2
1 bar/14.5 psi optimum (0.8–1.2 bar / 11.6–17.4 psi)
*
only for the CO
2
/O
2
version
CO
2
Control
Control range
0-15%
Accuracy
±(0.2%vol. + 3% of reading) e.g. ±0.35% at 5% CO
2
concen-
tration
Resolution
0.1%
Self-calibration
At least once per week the Gas Mixer must be either turned
off for five minutes or the CO
2
concentration must be set to
0% for five minutes This is necessary for the sensor’s on-
board automatic calibration routine, which needs an atmo-
spheric CO
2
measurement to combat sensor drift.
O
2
Control
∗
Control range
1-21% (Note: Settings below 1% are less accurate and a typi-
cal minimum value in an ibidi Heating Chamber is 0.5% O
2
)
Accuracy
±0.2% (typical)
±0.5% (max.)
Resolution
0.1%
Heat-up time
Five minutes before proper operation of sensor
Version 2.2 (May 15, 2019)
8
Instruction Manual