ibidi Heating System, Multi-Well Plates, K-Frame
5
Operation
Before starting an experiment, check that the ibidi Heating System, Multi-Well Plates, K-Frame fits on
your microscope stage, and that your multi-well plates are compatible to the Heated Plate. Connect
all parts (Section
) and perform a temperature adjustment as explained in Section
Important!
The Temperature Controller only controls the temperature at the internal sensors placed on the
Heated Lid and in the Heated Plate. The resulting temperature in the sample must be adjusted
for your specific setup. Follow the instructions in Section
5.1
Installation and Connection of the Parts
The components of the ibidi Heating System, Multi-Well Plates, K-Frame are connected as shown in
Figure
Figure 9 – Installation of the components of the ibidi Heating System, Multi-Well Plates, K-Frame.
1. Place the Temperature Controller next to the microscope and connect the power supply.
2. Confirm that the power switch of the Temperature Controller is off.
3. Insert the Heated Plate into the opening of the microscope stage.
4. Place the Heated Lid onto the Heated Plate.