Manual
ibaPADU-8AI-U/-8AI-I
18
Issue 1.6
7.3.6
RJ11 connection X9
A notebook equipped with an ibaCom-PCMCIA-F card can be connected at the RJ11
jack at the bottom of the device.
Thus, it is possible to carry out measurements in parallel at the RJ11 jack without affect-
ing the data transmission on the fiber optic cable.
7.4
Operating elements
7.4.1
Rotary switch S1
S1 is used to set the device address (1 … 8). Each address represents 8 A + 8 D chan-
nels.
Important note
The position of the address switch S1 must not be “0, 9…F”, since otherwise the link is
broken.
7.4.2
Rotary switch S2
7.4.2.1 ibaPADU-8AI-U
S2 is used to set the device mode (different measuring range, filters, input impedance):
Mode (S2)
Measuring
range
Filter*
input im-
pedance
Replacement for device
0
±10 V
1
100 k
Ω
ibaPADU-8
1
±10 V
1, 2
100 k
Ω
(new)
2
±10 V
1, 2, 3
100 k
Ω
ibaPADU-8-F1
3
±10 V
1
1 M
Ω
ibaPADU-8-HI
4
±10 V
1, 2
1 M
Ω
(new)
5
±10 V
1, 2, 3
1 M
Ω
ibaPADU-8-HI-F1
6
±24 V
1
100 k
Ω
(new)
7
±24 V
1, 2
100 k
Ω
(new)
8
±24 V
1
1 M
Ω
ibaPADU-8-HI-25
9
±24 V
1, 2
1 M
Ω
(new)
A
±60 V
1
100 k
Ω
ibaPADU-8-60
B
±60 V
1, 2
100 k
Ω
(new)
C
±60 V
1
1 M
Ω
ibaPADU-8-HI-60
D
±60 V
1, 2
1 M
Ω
(new)
E, F
undefined
*Filter: 1 = R/C low-pass 4 kHz, 2 = Antialiasing digital 330 Hz, 3 = Butterworth, digital 250 Hz