Guidance and manufacturer’s declaration – ELECTROMAGNETIC IMMUNITY – 

FOR EQUIPMENT AND SYSTEMS THAT ARE NOT LIFE-SUPPORTING

The device is intended for use in the electromagnetic environment specified below. 

The customer or the user of the device should assure that it is used in such an environment.

Immunity test

Test level

IEC 60601

Conformity level

Electromagnetic 

environment

- guide

Portable and mobile RF communications equipment should not be used near any part of the device, 

including cables, except when the recommended separation distance is respected, calculated from 

the equation applicable to the frequency of the transmitter.

Recommended separation distance

Conducted RF

IEC 61000-4-6

3V

eff

 from 150kHz

to 80MHz

3V

eff

 from 150kHz to 

80MHz

d = 1,2 √P

from 150kHz to 80MHz

RF irradiate

IEC 61000-4-3

10V/m from 80MHz 

to 2.7GHz

10V/m from 80MHz 

to 2.7GHz

d = 2,3 √P

from 80MHz to 800MHz

d = 1,2 √P

from 800MHz to 2,7 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter 

manufacturer and d is the recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveya, should be 

less than the compliance level in each frequency rangeb.

Interference may occur in the vicinity of equipment marked with the symbol above:

         

Note:

(1)  At 80 MHz and 800 MHz, the higher frequency range applies.

(2) Electromagnetic propagation is affected by absorption and reflection from structures,

  objects and people.

a)  Field strengths from fixed RF transmitters, such as base stations for radio (cellular/cordless)

  telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 

  cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to

  fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field

  strength in the location in which RESTART is used exceeds the applicable RF compliance level above,

  RESTART should be observed to verify normal operation. If abnormal performance is observed,

  additional measures may be necessary, such as re-orienting or relocating RESTART.

b)  Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.

Содержание WEPERE RESTART

Страница 1: ...USER MANUAL Recover faster anytime anywhere...

Страница 2: ......

Страница 3: ...e description and controls 7 Labelling 8 Pack contents 10 How To Use Introduction to the technology 10 Contraindications 11 Warnings 12 Device use 12 Treatments 14 Looking after the device Maintenance...

Страница 4: ...ns of the electromagnetic compatibility Directive 2014 30 UE of the European Parliament and of the Council of the 26th February 2014 to the current TECHNICAL STANDARD on ELECTROMAGNETIC COMPATIBILITY...

Страница 5: ...able anaesthetic mixture with air or with oxygen or with nitrous oxide device intended for continuous operation device not suitable for external use Intended purpose and scope of use The RESTART devic...

Страница 6: ...0Hz 10 Duty cycle 5 50 100 mains powered 5 50 100 battery powered Power Can be set in 3 steps L low M medium H high Maximum power density 1 6W cm mains powered 0 8W cm battery powered Maximum output p...

Страница 7: ...L Device description and controls 1 Treatment time indicator 2 Time button 3 Power adjustment button 4 Treatment power indicator 5 Battery indicator 6 Device ON OFF button 7 Emitting head 8 Power supp...

Страница 8: ...Out 15V 1 2A Battery Ni MH AAA850mAh 4 8 V Ultrasound Carrier frequency of use 1MHz 10 Duty cicle 5 50 100 mains powered 5 50 100 battery powered Maximum output power 6 4W mains powered 3 2W batter po...

Страница 9: ...documents accompanying the product IP22 Device protected against the penetration of solids with diameter d 12 5mm and against the fall of vertical drops of water when the device is held at 15 from the...

Страница 10: ...sound waves would have to over come to reach the body To reduce this difference a coupling substance must be used If there is a small amount of air between the transducer and the skin the proportion o...

Страница 11: ...advised not to wear metal objects to use ONLY the accessories supplied by the manufacturer It is forbidden to use the device in the presence of equipment for monitoring the user s vital functions equi...

Страница 12: ...s used in strict compliance with the instructions given in this manual Should any foreign substances get into the device contact the dealer or manufacturer immedia tely Should the device fall check th...

Страница 13: ...has been set and the head has been placed in contact with the skin the treatment will begin it is important to perform the treatment by continuously and uniformly moving the head around the ap plicat...

Страница 14: ...manufacturer Treatment Intensity Minutes Muscle H Power supply Battery 15 Tendon Ligament L M Power supply M Battery 10 15 Aesthetic L Power supply L M Battery 5 10 REMEMBER TO keep the ultrasound hea...

Страница 15: ...elative humidity from 15 to 93 pressure from 700 to 1060 hPa In the bag provided ambient temperature from 5 to 40 C relative humidity from 15 to 93 pressure from 700 to 1060 hPa Troubleshooting Any ty...

Страница 16: ...ry indicator will flash from L M H in sequence Once char ging is complete the battery charge indicator will light up as H High WARNING the battery life cycle depends on the charge discharge cycles to...

Страница 17: ...ided by IACER Shouldyouneedtoreturnthegoodsthenpleasepackthedeviceandalltheaccessoriessothatitwon t be damaged during transportation In order to be entitled to the warranty assistance the purchaser mu...

Страница 18: ...facturer if non original spare parts are used the operation and safety of the product might be affected and the warranty will be null and void Interference and electromagnetic compatibility tables The...

Страница 19: ...ns and to use caution in therapeutic applications on users with pacemakers However it is advisable to use the device keeping a distance of at least 3 metres from televisions monitors mobile phones or...

Страница 20: ...USER MANUAL 20 RESTART All rights reserved RESTART and the logo are the exclusive property of I A C E R S r l and are registered trademarks Issue MNPG417 01 of 31 01 2022...

Страница 21: ...21 USER MANUAL...

Страница 22: ...omagnetic environment guidance RF emissions CISPR 11 Group 1 The RESTART uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interfere...

Страница 23: ...line 2kV Line s to line 1kV Line s to line 2kV Line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variation...

Страница 24: ...nsmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compl...

Страница 25: ...ation distance according to the frequency of the transmitter m d 1 2 P from 150kHz to 80MHz d 1 2 P from 80MHz to 800MHz d 2 3 P from 800MHz to 2 7 GHz 0 01 0 12 0 12 0 23 0 01 0 38 0 38 0 73 1 1 2 1...

Страница 26: ...USER MANUAL 26...

Страница 27: ...27 USER MANUAL...

Страница 28: ...I A C E R Srl via Enzo Ferrari 2 30037 Scorz VE Italy www wepere com...

Отзывы: