USER MANUAL
14
N.B.
disconnect the cables before putting the device in the bag. If this is not done, the cables can
be excessively bent near the connectors, which can damage the cables.
WARNING:
to ensure user safety, the device is equipped with a system that detects cor-
rect coupling between the ultrasound head and the user’s skin.
In the event of incorrect
coupling or bad contact, the treatment time LED will start flashing and the ultra-
sound intensity will be reduced
. Once head/skin contact is restored, the intensity will
automatically increase slowly, up to the previously set level.
WARNING:
to ensure user safety, the device is also equipped with a temperature regula-
tion system.
If the head temperature exceeds 42°C, the device will end the treatment
and the time indicator LED will flash twice
; it will not be possible to resume treatment
until the head reaches a temperature below 40°C.
Treatments
Below is the list of treatments suggested by the manufacturer:
Treatment
Intensity
Minutes
Muscle
H - Power supply/Battery
15
Tendon/Ligament
L/M - Power supply M - Battery
10 – 15
Aesthetic
L - Power supply L/M - Battery
5 – 10
REMEMBER TO:
•
keep the ultrasound head always in motion;
• use a good amount of gel to ensure contact;
• act evenly on the treated area.
Maintenance
If used in accordance with the information reported herein, this device requires no particular rou-
tine maintenance operations.
It is advisable to carry out a functional check of the device at the Manufacturer every 24 months.
The manufacturer does not consider the RESTART device to be repairable by personnel outside
the company itself. Any such operation by personnel not authorised by the Manufacturer will be
considered tampering with the device, thereby avoiding the manufacturer’s warranty and freeing
it from liability for any hazards to which the operator or user may be subjected.
CLEANING
It is advisable to switch off the RESTART at the end of each treatment session, in addition to
removing the power cable.Use a soft dry cloth to remove any dust from the device. In case of
hard-to-remove dirt, use a cloth soaked in water and alcohol.