4. Turn on the device by moving the switch to the ON position. The ultrasound intensity indicator
will display L (Low, preset), whereas the battery level indicator will indicate the battery capacity:
low (L-Low preset), medium (M-Medium) and high (H-High).
5. Select the desired intensity by repeatedly pressing the
MODE
key. There are three levels of in-
tensity selectable in order: low (preset L – Low), medium (M – Medium) and high (H – High).
6. Select the treatment time by repeatedly pressing the
TIME
key: the LEDs relating to the 5-10-15
minutes of treatment will light up in sequence, as shown in the figure. When in use, the chosen
treatment time indicator will be constantly lit until the set time has elapsed.
7. Once the application time has been set and the head has been placed in
contact with the skin, the treatment will begin: it is important to perform the
treatment
by continuously and uniformly moving the head around the ap-
plication area, with slow and circular movements
. The treated area should
be twice the applicator diameter. If there is poor transmission of ultrasound
energy, it is recommended to add more gel or reposition the ultrasound head.
Head movement should not be too slow to avoid inducing heat; nor too fast to pre-
vent bad contact which would reduce the effectiveness of the treatment.
8. At the end of the treatment all the indicator lights will turn off. Turn the switch to
OFF
and di-
sconnect the device from the power supply (not needed if you are using the device with bat-
tery).
9, Clean any gel off the head before storing the device and its accessories in the bag provided.
Make sure there is no gel left on the head. DO NOT SUBMERGE IN WATER!