Guidance and manufacturer’s declaration
ELECTROMAGNETIC IMMUNITY – FOR ALL EQUIPMENT AND SYSTEMS
The RESTART is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment
Immunity test
Test level
IEC 60601
Compliance level
Electromagnetic
environment - guide
Electrostatic
discharge (ESD)
IEC 61000-4-2
8kV contact
15kV on air
8kV contact
15kV on air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2kV for power
supplies lines
2kV for power
supplies lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Overvoltage
IEC 61000-4-5
1kV Line(s) to line
2kV Line(s) to line
1kV Line(s) to line
2kV Line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips,
short interruptions
and voltage variations on
power supplì input
lines
IEC 61000-4-11
<5% U
T
(>95%dip in U
T
)
for 0,5 cycles
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
25 cycles
<5% U
T
(>95% dip in U
T
)
for 5s
<5% U
T
(>95% dip in U
T
)
for 0,5 cycles
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5s
Main power quality
should be that of a typical
commercial or
hospital environment.
If the user of RESTART
requires continued
operation during
power mains
interruptions, it is
recommended that
MIO_SONIC
be powered from an
uninterruptible power
supply (UPS) or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30A/m
30A/m
Power frequency
magnetic fields should be
at levels characteristic
of a typical location in a
typical commercial
or hospital
environment
Note: UT is the A.C. mains voltage prior to application of the test level.