Step 8

Obtain a blood sample using the 

lancing device. Place the device 

against the pad of the finger. The best 

puncture sites are on the middle or 

ring fingers. Press the release button. 

Remove the device from the finger. 

Wait a few seconds for a blood drop to 

form. A minimum volume of 0.5 microliter is needed to fill the 

confirmation window. (actual size of 0.5 

 :   )

Step 9

After the 

 symbol appears on 

the screen, apply the blood sample 

to the narrow end of the test strip till 

the meter beeps. If the confirmation 

window is not filled before the meter 

finishes counting down then discard the 

test strip and insert a new one.

If confirmation window is not filled in 

time because of abnormal viscosity or 

insufficient volume, Er4 message will 

appear.

Step 6

Insert a test strip with the contact bars 

facing upwards into the meter’s test 

strip port. Push the strip in gently until 

the meter beeps. Be careful not to bend 

the test strip. The 

 symbol will 

appear on the screen.

Step 7

 

If you want to attach an post-meal flag ( ) to 

a test result, press and hold   for 3 seconds 

after inserting the test strip. The post-meal flag 

( ) and the 

 symbol will appear on the 

screen.

The test result will also be displayed with the 

post-meal flag ( ).

If you do not want to save the result as an 

postmeal test, move on to the step 8 after the step 6.

Preparing the Meter and Test Strip

Flagging Post-meal Test Results

The CareSens N meter allows you to flag a result of an 

postmeal test with ( ) icon. The post-meal test flag ( ) can 

be attached just before applying the blood sample. Once you  

attach the post-meal flag ( ) to the test results, it cannot be 

deleted.

Applying Blood Sample

Good Sample

Insufficient 

Sample

Note:

 

The meter may switch off if the blood sample is not applied 

within 2 minutes of the 

 icon appearing on the screen.

If the meter turns off, remove the strip and reinsert after 

switching on the meter.

Содержание CareSens N Voice

Страница 1: ...c All Rights Reserved PGA1E3121 REVx 02 2012 The Biosensor Exppert i SENS Inc 465 6 Wolgye4 dong Nowon gu Seoul 139 845 Korea www i sens com Medical Technology Promedt Consulting GmbH Altenhofstrasse...

Страница 2: ...stem The system provides you with safe convenient and painless blood glucose in vitro i e outside the body diagnostic monitoring You can obtain accurate results in just 5 seconds with a small 0 5 bloo...

Страница 3: ...Lancing Device 21 Preparing the Lancing Device for Blood Sample Retrieval 22 Preparing the Meter and Test Strip 24 Flagging Post meal Test Results 24 Applying Blood Sample 25 Discarding Used Lancets 2...

Страница 4: ...reSens N strip The average hematocrit range is between 30 and 50 However people with severe anemia and newborns may have higher hematocrit and this can affect the test results Very high 60 or more or...

Страница 5: ...C 50 104 F Relative humidity 10 90 Hematocrit 20 60 CareSensNBloodGlucoseMonitoringSystem Components 01 CareSens N Blood Glucose Meter 02 Lancing Device 03 Lancets 10 04 Owner s Booklet 05 Quick Refe...

Страница 6: ...ps past the expiration date Test strips in new unopened containers and test strips in containers that have been opened can be used up until the expiration date printed on the test strip box and contai...

Страница 7: ...lts stored in the memory are displayed alarm appears when the time alarm has been set mmol L unit for measuring blood glucose Battery icon indicates meter battery is running low and needs to be replac...

Страница 8: ...adjust until the correct year appears Press and hold button to scroll through the numbers quickly After setting the year press the S button to confirm your selection and enter the month setting mode S...

Страница 9: ...is case the unit can not be changed and the unit change setting step will be excluded Step 9 On pressing or the screen will display the On or OFF Press the S button to confirm the selection The meter...

Страница 10: ...e and is used to check that the meter and the test strips are working properly The test strip vials have CareSens Control Solution ranges printed on their labels Compare the result displayed on the me...

Страница 11: ...displayed on the screen Step 3 A test result will appear after the meter counts down from 5 to 1 After your control solution result appears on the display press for 3 seconds till the check icon appe...

Страница 12: ...e contact your authorized i SENS sales representative Caution To avoid infection when drawing a sample use a lancet only one time and Do not use a lancet that has been used by others Always use a new...

Страница 13: ...Sample Retrieval Step 4 Select a desired depth of one to five 1 5 on lance s adjustable tip Rotate ring to align desired number with arrow A beginning setting of three 3 is recommended Step 5 To cock...

Страница 14: ...meter s test strip port Push the strip in gently until the meter beeps Be careful not to bend the test strip The symbol will appear on the screen Step 7 If you want to attach an post meal flag to a t...

Страница 15: ...ically switch off after 3 seconds Discard used test strips safely in disposable containers Step 1 Unscrew general lancing device tip Step 2 Place protective cover on lancet Push the lancet ejector for...

Страница 16: ...observed during testing AlternativeSitesforTesting Things to know when using AST Please understand the following things before testing outside of the fingertip arms palms thighs and calves The capilla...

Страница 17: ...splayed even though you do not have hyperglycemia or hypoglycemia AST Precautions Do not ignore the symptoms of hyperglycemia or hypoglycemia When the results of the test do not reflect one s opinion...

Страница 18: ...the last 14 days ViewingTest Results Stored in the Meter s Memory Step 2 Press to view the average value and the number of tests performed before eating a meal for the last 14 days On pressing again t...

Страница 19: ...will be displayed The screen will then automatically change to the memory check mode At this time the bell icon indicating that the PP2 alarm has been set will be displayed on the screen Step 2 Setti...

Страница 20: ...ncel the alarm1 The OFF icon will blink on the screen Step 3 Press to adjust the time of the alarm1 A number representing the time will blink on the screen Press to set the time Press to end Setting t...

Страница 21: ...nto place Note Removing the meter batteries will not affect your stored result However you may need to reset your meter settings See pages 12 14 Caring for Your System Use a soft cloth or tissue to wi...

Страница 22: ...40 C 50 104 F and repeat the test after 30 minutes Understanding Error and Other Messages Note If the error messages persist contact your authorized i SENS sales representative Message What It Means...

Страница 23: ...r authorized i SENS sales representative Performance Characteristics The performance of CareSens N Blood Glucose Monitoring System Strips has been evaluated in laboratory and in clinical tests Accurac...

Страница 24: ...owsthattherecouldbevariationofupto4 2 Warranty Information Manufacturer sWarranty i SENS Inc warrants that the CareSens N Meter shall be free of defects in material and workmanship in normal use for a...

Отзывы: