Hyquest Solutions Pty Ltd

 

Top Set Rod – Model: MagnaRod 

© Copyright 

 

 

 

 

- 22 -

   

 

Issue 1.03   5 Mar, 2013 

The  following  information  details  some  specifics  about  each  format,  specifically  the 
limitations and alternate characters that may be presented. 

1)  

The maximum time that can be displayed, during the measurement, is ‘FFFF

16

’.  If the 

measurement  continues  beyond  the  maximum  data  form  available,  the  count  ‘rolls 
over’ and begins from ‘0’.  When this occurs, the delimiter character ‘,’ is replaced by 
the ‘>’ character to indicate the rollover condition.  When the data is transferred to a 
spreadsheet,  the  character,  ‘>’,  can  be  easily  detected  and  used  to  add  ‘65535’  (= 
FFFF

16

) to the elapse time tally.  No further change in the delimiter character occurs if 

the accumulating time continues to further rollover. 

2) 

The  maximum  time  that  can  be  displayed  as  given  by  the  final  data  string  is  436.4 
seconds as displayed (FFFF

16

 x 0.00666 seconds).  This equates to 7 minutes and 16.4 

seconds.    If  the  accumulating  time  continues  during  the  measurement  (as  discussed 
above), the final  count  tally will reflect  a calculated value rolled over and beginning 
from ‘0’.  If this occurs, the delimiter character ‘,’ is replaced by the ‘>’ character to 
indicate  a  rollover.    To  determine  the  total  accumulated  time  from  the  roll  over 
condition, simply add 436.4 seconds to the total number displayed. 

3) 

The  maximum  number  of  measured  meter  contact  closures  (clicks)  is  999.    If  the 
accumulating  contact  closures  continues  beyond  the  maximum  data  form,  the  count 
‘rolls  over’  and  the  most  significant  digit  position  number  is  replaced  by  the  colon 
(‘:’)  character.    If  the  count  continues,  other  characters  will  replace  the  most 
significant  digit  position  number.  Meter  counts  should  not  exceed  999,  however  a 
detection scheme is provided if such an occurrence happens.  

 

The “Spin Test” has three subcommands as follows: 

 

S   –   Start spin measurement. 

 

A   –   Stop spin measurement, transmit final results, then quit. 

 

 I   –   Abort or quit the “Spin Test” function.  No final data is transmitted. 

 

The “Spin Test” is performed in the following manner (through the serial interface) : 

 

1.  Before  entering  the  “Spin  Test”  function,  select  the  type  of  meter,  i.e.  Cat’s 

Whisker Head, or Magnetic Head Meter.  Also enable or disable the buzzer. 

 

2.  Press  “

N

”  on  the  computer  keyboard  to  enter  the  “Spin  Test”.    The 

COUNT

 

LED  will  become  illuminated.    The 

OPERATING  STATUS

  indicator  will 

illuminate 

green

  if  the  meter  is  selected  for  a  Magnetic  Head  and 

red

  if  Cat’s 

Whisker Head is selected.  The 

MagnaRod 

echoes with ‘

N

’. 

 

3.  Depress the letter “

S

” on the computer keyboard or press the SELECT button on 

the 

MagnaRod

.   The 

OPERATING STATUS

 indicator will turn off.  

 

4.  When the meter is spun, the 

OPERATING STATUS

 indicator will blink when 

a contact closure is detected. 

 

5.  The 

MagnaRod

  will  output  through  the  serial  port  a  character  string  (as 

Содержание MagnaRod MAGD/CMC

Страница 1: ...C HYQUEST SOLUTIONS PTY LTD PO BOX 332 LIVERPOOL B C NSW 1871 AUSTRALIA Phone Int 612 9601 2022 Fax Int 612 9602 6971 Phone Nat 02 9601 2022 Fax Nat 02 9602 6971 Email sales hyquestsolutions com au We...

Страница 2: ...4 Current Meter Counter Operation 8 4 5 Current Meter Counter Menus 9 4 6 Indicator LEDs 12 4 7 Bluetooth Communications 12 4 7 1 Serial Communications Interface 13 4 7 2 Data Conversion Example 15 4...

Страница 3: ...quations and 3 rotations sec ranges The MagnaRod counter operates with 3 x AA alkaline 1 5V batteries and comes in a robust waterproof enclosure The commands are identical to the CMCsp commands so exi...

Страница 4: ...nal to determine if a contact closure has occurred and if successive signals are of the same characteristics The MagnaRod then determines contact closure conditions and either counts the signal closur...

Страница 5: ...ress and hold both buttons to turn on the MagnaRod Power OFF Press and hold both buttons again to turn off the MagnaRod Left Right Button Button Bed TSR Surface Stop CMC Start Up Dn Menu Select Hold T...

Страница 6: ...being measured from the surface down to the current meter The total water depth or Sounding is displayed on the bottom of the screen Depth too small indicates you have moved the rod less than 0 02ft 5...

Страница 7: ...e from the river bed to the current meter shaft when the Top Set Rod is in the lowest position 3 LCD Options allows 3 more LCD selections to be chosen The contrast can vary between 6 light and 24 dark...

Страница 8: ...s pulses are detected 3 When the elapsed timer exceeds the measurement interval and a meter pulse is detected the timer will be stopped The final count elapsed time and the calculated velocity are all...

Страница 9: ...contacts this becomes extremely important since Cat s Whisker contacts offer the poorest quality for marking and counting rotations In Cat s Whisker mode there are 2 speed settings FAST is selected fo...

Страница 10: ...tions on the screen match the equations on your Current Meter calibrations sheets 6 Select Save and Return to go back to the previous menu 7 When More Settings Test is selected 5 more selections becom...

Страница 11: ...e for the magnetic head contact does not apply but can be performed 11 The Spin Test is an elapsed time measurement from the time a meter is spun to when rotation of the meter stops Press Start and th...

Страница 12: ...d IEEE 802 14 that offers a transparent serial port protocol for communication When the MagnaRod is turned On the MagnaRod determines if Bluetooth is installed If available Bluetooth can be enabled an...

Страница 13: ...n such as HyperTerm Commands can be sent to the MagnaRod to perform special functions or select operating modes The following describes the serial data requirements commands and data formats that are...

Страница 14: ...t that is transmitted to the externally connected device can be used to display elapsed time and velocity while the measurement is being performed At the end of the measurement interval or if the T co...

Страница 15: ...uously output a 1 second data display without a time limit The T or I command must be used to terminate the operation V The software version of the MagnaRod is sent in the following manner vX X Exampl...

Страница 16: ...terval time is then used to compute the measured velocity For the Normal mode the 1 second output data format displays the accumulating time interval counts based on a 300 Hz sample rate The following...

Страница 17: ...interface Command Function Performed N Spin Test S Start Spin Measurement subcommand in Spin Test A Stop Spin Measurement subcommand in Spin Test U Buzzer ON Z Buzzer OFF M Magnetic Head Meter C Cat...

Страница 18: ...ove the batteries and any remaining water Allow the unit to dry out completely before reinstalling the batteries There are no user serviceable parts inside the MagnaRod If the unit does not function p...

Страница 19: ...porary immersion Note 60 1 Meter cup width for 6 cups Water Velocity Flow absolute maximum 1 Normal Mode Price AA meter Mag 45 ft s 13 7m s 99o to 204o dwell CatW 22 ft s 6 7m s 10o to 62o dwell Pygmy...

Страница 20: ...er Head 1 NOTES a Measurements and Computations based on the following Rating Table Equations Price AA Meter V 2 2048 0 0178 where V Velocity ft s T T Seconds rotation Pygmy Meter V 0 9604 0 0312 T b...

Страница 21: ...f the spin test the MagnaRod sends a final data string that displays the total number of counts and total time all shown in decimal notation This format makes it easy to copy the final total number of...

Страница 22: ...nds to the total number displayed 3 The maximum number of measured meter contact closures clicks is 999 If the accumulating contact closures continues beyond the maximum data form the count rolls over...

Страница 23: ...in the Accessories folder listed in the Programs list on the desktop Within HyperTerminal and before spinning the meter select Capture Text found in the Windows label Transfer tab When the Capture Tex...

Страница 24: ...Do not import column skip in the Column data format window This will show Skip in the column 1 heading 11 Click on column 3 where the Time data is shown in HEX ASCII and change in the Column data form...

Страница 25: ...response of a current meter created from data in a Spin Test file Spin Test for Meter 1234567 0 10 20 30 40 50 60 70 80 90 100 110 120 130 1 6 11 16 21 26 31 36 41 46 51 56 61 66 71 76 81 86 91 96 101...

Страница 26: ...wing simple formula Dwell Angle Contact time x 360 Total Time The contact time and total time shown above in the formula are expressed in terms of the number of samples measured at a sample rate of 12...

Страница 27: ...e 27 4 06 INSPECTED __________________ Anthony Gemayel Mechatronics Engineer _____________________________________________________________ The above meter has been calibrated in accordance with AS 377...

Отзывы: