Configuration

§

Echotest

: Clicking  this  button,  the  signal  strength  of  the  433  MHz  near  field  radio  connection  will  be  tested.  Data

packages will be sent, until the test is aborted manually by the user.

§

Update Logger Firmware

: Perform an update of the logger’s firmware. After a security prompt, new firmware files can

be selected a transferred to the logger. Please refer to the chapter “Updates” for further information.

§

Update  USB-Dongle  Firmware

:  Perform  an  update  of  the  USB-dongles  firmware.  After  a  security  prompt,  new

firmware files can be selected and transferred to the USB-dongle.  Please  refer  to  the  chapter  “Updates”  for  further
information.

§

All Radio OFF

: All loggers in range of the dongle will be set to 'flight mode', so the they will not send any more data.

Important

: The loggers will also disappear from the list in GP-Shell.

§

All Radio ON

: All loggers in range of the dongle will be set from 'flight mode' back to 'normal mode'.

3.1.3

Main Window

§

Identify Logger

: Establishes a connection to a logger. To connect to a logger, select a logger from the list before clicking

“Identify Logger”.

§

Synchronize Clock

: Synchronizes logger time with time of the current system.

§

Display Measure

: Show current measurement values (status of logger and measured values of the activated measuring

probe). The values measured when clicking this button will not be stored and not be transmitted to the server, but can
be used for local calibration of data loggers.

§

Load to  Disk

:  Download  measured  values  from  the  data  logger  to  your  local  computer.  You  can  choose  between

incremental download (only values that have not been  transmitted  to  the  server  yet  will  be  downloaded)  and  full
download (all values stored on the logger will be downloaded). The local path for saving the data  can  be  set  in  the
setup menu (see 

Setup

).

§

Attention

: If the modem of the logger is switched off, any data downloaded to your computer will be marked as sent

and will not be transmitted during the next GPRS-transmission. In doing so, the loggers can be operated even without
the use of the GPRS-connection.

§

Clear  Logger

:  By  clicking  this  button,  the  whole  memory  of  the  logger  will  be  cleared.  (All  measured  data  will  be

deleted!)

§

Parameters

: Opens a menu for configuring a logger. The logger must be identified first.

§

Data from Disk

: Open Files that have been downloaded to the local  computer  previously,  i.e.  measured  values  (i.e.

*.g2p files).

§

Exit

: Closes the application.

10

Содержание iLevel-GW

Страница 1: ...iLevel GW User Manual...

Страница 2: ...8 Part III Configuration 9 3 1 Functions Settings 9 3 1 1 Setup 10 3 1 2 Service 10 3 1 3 Main Window 11 3 1 4 Parameters 12 3 1 5 Scale 13 3 1 6 More Buttons 13 3 2 GPRS GSM Setup 14 3 3 Two Point Ca...

Страница 3: ...rmation provided in this manual was deemed accurate as of the publication date However updates to this information may have occurred This manual does not include all of the details of design productio...

Страница 4: ...und before starting to work as well as for defects found while working Do not use the HyQuest Solutions iLevel GW in areas where there is a danger of explosion The present user manual specifies enviro...

Страница 5: ...nd more specifically to determine groundwater and surface water levels Never use the the HyQuest Solutions iLevel GW to transport people Special caution is needed for all works and operations in the v...

Страница 6: ...GW comes with a coupled water level and water temperature sensor accurate to 0 1 FS The vented pressure sensors auto compensate for atmospheric pressure and temperature variations providing a correcte...

Страница 7: ...underfloor installations underfloor antennas are available Those are clamped between the frame and the lid Helical antennas are fitted onto the lid for above floor measuring points For fitting the ant...

Страница 8: ...rting the battery incorrectly may blow the fuse which can only be repaired at the factory or even damage the electronics The electronic comes with a backup capacitor which can compensate a power outag...

Страница 9: ...needed for the installation of the dongle are installed with the HyQuest Solutions GP shell Please use the supplied CD or download the latest version of the software from the customer homepage Double...

Страница 10: ...your export file Optional Internet Upload Enabling this option data from third party loggers can be transmitted to the server Language Language for the user interface can be chosen from a drop down me...

Страница 11: ...ronize Clock Synchronizes logger time with time of the current system Display Measure Show current measurement values status of logger and measured values of the activated measuring probe The values m...

Страница 12: ...next modular time i e at a period of 1 h this would be the next full hour If for example a period of 12 hours is selected the measurements would be made at 0 00 and 12 00 With the Period Offset the ti...

Страница 13: ...larms Check this option if you want to set special alarm events e g shorter measuring and transmissions intervals 3 1 5 Scale Offset Offset available if no pressure sensor is used for the water level...

Страница 14: ...nding connection errors GSM period Set the transmission intervals in which the data is transmitted The minimum for this setting is five minutes Use the drop down menu to switch between seconds minutes...

Страница 15: ...r will delete all logged data when 95 of the memory is used This will happen only if all data has been transmitted successfully GPRS Offset sec Time difference between the logger and the server in sec...

Страница 16: ...ing the button Measure Set Point Nominal Enter the required current value This can be a currently measured tapping or the water level taking the main sea level into account Calculate and OK If all set...

Страница 17: ...The software HyQuest Solutions GP Shell does not require any maintenance by the customer If an updated version is released 4 2 Loggers Our loggers only require very low maintenance Do not use any shar...

Страница 18: ...your return shipment 2 Please provide as much information and or clear instructions within the return paperwork This will assist our test engineers with their diagnosis 3 Please do not ship the goods...

Страница 19: ...finable minimum 2 Material and Environmental Conditions Enclosure stainless steel ABS Operating temperature range Modem 30 C to 85 C Logger 40 C to 85 C Dimensions Diameter Length Mass 48 5 mm 411 5 m...

Страница 20: ...ctions must be observed 7 2 Dismantling Disposal When disposing of the units and their accessories the applicable local regulations regarding environment disposal and occupational safety must be obser...

Страница 21: ...re committed to supporting efforts to avoid and recycle waste Please consider Avoidance before recycling Recycling before disposal This symbol indicates that the scrapping of the unit must be carried...

Страница 22: ...ide the device rises rapidly within a very short time the hermetic sealing might be damaged This can happen due to material expansion caused by big temperature difference Please replace O rings and ca...

Страница 23: ...S Group HyQuest Solutions KISTERS LATAM HyQuest Solutions NZ Ltd HyQuest Solutions Pty Ltd HyQuest Solutions KISTERS AG 49 2408 9385 0 info hyquestsolutions eu 612 9601 2022 sales hyquestsolutions com...

Страница 24: ...www hyquestsolutions com 2022 HyQuest Solutions 21 04 2022...

Отзывы: