Firmware Version 1.2
MultiControl 4070
4
Safety
ENG
Handling information for the MultiControl 4070
•
Never expose the instrument to excessive heat, cold, or humidity.
•
Do not store the instrument in a humid or dusty location.
•
Never submerge the instrument into water or other liquids. Never let
liquids come into the instrument.
•
Never open the instrument.
•
Do not use the instrument if it has been dropped or if the casing is
damaged.
•
Avoid strong magnetic fields. Keep the instrument away from electric
motors or other instruments which generate electromagnetic fields. Strong
magnetic fields may cause malfunctions and influence measurement
values.
•
Prevent the formation of condensation. If condensation has formed, let the
instrument acclimate before you switch it on.
Information about the use of sensors and cables
•
Protect the sensors from exceeding the allowed power range, mechanical
overload and incorrect pin assignment.
•
Make sure you enter the sensor parameters correctly in the HYDROlink6
software when using sensors without an ISDS (Intelligent Sensor Detec-
tion System).
•
When extending standard measuring cables, use the extension cables
provided for this purpose, otherwise the shielding will be interrupted.
•
The data of ISDS sensors are read in when the measuring instrument is
switched on. If ISDS sensors are reconnected, the measuring instrument
must be switched off and on again for the sensor data to be adopted.
•
Never connect an inductive load to the digital signal input.