![Hydrotechnik HySense QL 326 Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/hydrotechnik/hysense-ql-326/hysense-ql-326_manual_2192491036.webp)
Calibrate end value of channel 1
ID
DLC
B0
B1
B2
B3
B4
B5
B6
B7
0x620
8
77h
02h
01
00
00
00
00
00
The D/A output is set to 0 V. To get the actual digital value for the final value, repeat the message until the DVM
displays 10 V DC at the terminal output. With the following message the AD value is decremented again so that
the exact final value can be determined.
ID
DLC
B0
B1
B2
B3
B4
B5
B6
B7
0x620
8
77h
02h
00
00
00
00
00
00
Once the final value has been set, confirm this with the message:
ID
DLC
B0
B1
B2
B3
B4
B5
B6
B7
0x620 8
77h
02h
03
00
00
00
00
00
The end value of channel 1 is now stored in the device.
Calibrate zero and full scale value of channel 2
Perform the same procedure for channel 2 as for channel 1. Replace 0x78 in the CAN messages for byte B0.
End calibration
Send this CAN message:
ID
DLC
B0
B1
B2
B3
B4
B5
B6
B7
0x620
8
99h
04h
00
00
00
00
00
00