Example for control via CAN
• Read QL Type ("QL326")
Command 0x55 0x01...
• Read the heartbeat status (Byte[0]) 0x02 means ControlUnit is not configured.
• Read firmware version
Command 0x55 0x02...
• Activate Send parameters ONCommand
0x99 0x01
• Send all parameters to control box (ramp and sine)
Param W1Param
Z1Param
W2Param
Z2Param
Sine
• Memory Parameters
Command 0x60
• Param Operating mode
Command 0x06
• Activate send parameters OFF
Command 0x99 0x02
HeartbeatState must now be sent 0x04.
• Send Connect to control box
Command 0x07 0x01HeartbeatState
0x05 must now be sent.
• Start test sequence
Command
NOTE: The control box now works with a timeout which is counted up internally. This timeout
must be reset at least every 1000ms via CAN command (0x2050/1), otherwise the control box
switches itself off and ends the test cycle.
Tx: 0x40 50 20 01 00 00 00 00
Rx: 0x43 50 20 01 05 00 00
HeartbeatState must now be sent 0x06.
Operating mode Ramp / Sine
The test cycle runs continuously until all cycles have been processed or the STOP command is
sent manually.
HeartbeatState must then be sent 0x05 again.
Manual operating mode
The desired valve value is sent.
Command 0x79
End test cycle
Command 0x08 0x02
Heartbeat state must be sent 0x05 again afterwards.