![Hydronix Hydro-Probe XT Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/hydronix/hydro-probe-xt/hydro-probe-xt_user-manual_2192332043.webp)
Hydro-Probe XT User Guide HD0538 Rev: 1.6.0 43
Chapter 5
Sensor Integration and Calibration
1 Sensor Integration
The sensor may be integrated into a process in one of three ways:-
•
The sensor may be configured to output a linear value of between 0-100 Unscaled units
with a material calibration being performed in an external control system.
Or
•
The sensor may be internally calibrated using the Hydro-Com sensor configuration and
calibration software to output an absolute moisture percentage value.
Or
•
The sensor could also be used as a target value
RS485 development tools are available from Hydronix for system designers who wish to develop
their own interface.
2 Sensor Calibration
2.1
Introduction to material calibration
The raw output of a Hydronix sensor is an Unscaled value ranging from 0 to 100 units. Each
sensor is set so that a 0 Unscaled value relates to the measurement in air and 100 relates to
water.
The relationship between the change in moisture % to the change in Unscaled value will not
be the same for two different materials as shown in Figure 34. This is because each material
has its own unique electrical characteristics. Calibration is the process of mapping the
sensors Unscaled readings to a value that represents the material moisture percentage.
20%
100
0
Material 1
B
Sensor unscaled reading
True moisture of material
Material 2
Calibration Line
C
B=
Moisture %
Unscaled Value
Figure 34: Calibrations for two different materials
Содержание Hydro-Probe XT
Страница 4: ...4 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 6: ...6 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 8: ...8 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 10: ...10 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 14: ...Chapter 1 Introduction 14 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 42: ...Chapter 4 Configuration 42 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 52: ...Chapter 6 Optimising Sensor and Process Performance 52 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 58: ...Chapter 7 Sensor diagnostics 58 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 60: ...Chapter 8 Technical specification 60 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...
Страница 66: ...Appendix B Document Cross Reference 66 Hydro Probe XT User Guide HD0538 Rev 1 6 0 ...